Chapter 9: Troubleshooting
9-4
Home Use Operation Manual
PRObleM
POSSIble CAUSe
CORReCTIVe ACTIOn
SpO
2
value is erratic,
intermittent, or
incorrect.
Poor patient perfusion.
Patient motion.
Defective sensor or patient
cable.
Confirm signal level with PI and signal
bar graphs.
Reposition the sensor on the patient.
Patient must remain still to obtain an
accurate measurement.
Try a new sensor or contact your
authorized service representative for
help.
Oximeter Alert.
Oximeter cable or sensor
is disconnected from the
oximeter.
Sensor is incorrectly
positioned on the patient.
Poor patient perfusion.
Oximeter searching too long
for initial SpO
2
signal.
Check sensor connections to the patient
cable and to the oximeter.
Reposition the sensor.
Confirm signal level with PI and signal
bar graphs. Reposition the sensor on the
patient.
Confirm signal level with PI and signal
bar graphs. Check sensor connections
to the patient cable and to the oximeter.
Reposition the sensor on the patient.
Printer and Data Communication
PRObleM
POSSIble CAUSe
CORReCTIVe ACTIOn
No printout on
optional printer.
AC power is not connected to
the Docking Station.
No trends in memory.
No paper or paper incorrectly
loaded.
Connect AC Power to the Docking
Station.
Take trend data.
See
Chapter 7: Optional Docking Station
and Printer
for details of loading paper.
“?” malfunction
indicator on the
printer is lit.
Printer interface malfunction.
Printer door open.
Printer out of paper.
Contact your authorized service
representative for help.
Close the printer door.
Install a roll of printer paper.
Real time or
trend data is not
transmitted.
An accessory cable is
defective.
Oximeter is not properly
seated in Docking Station.
The communications setup is
not correct.
Contact your authorized service
representative.
Remove the oximeter from the Docking
Station and replace it assuring proper
alignment.
Check the user-connected auxiliary
equipment and software.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......