Chapter 7: Optional Docking Station and Printer
Home Use Operation Manual
7-7
Choosing the Print Mode
Data can be printed in real time, numeric trend or graphic trend mode. In either trend mode,
up to 45 days of previously stored data (depending on trend interval) collected from 1 to 99
patients can be printed. The printer select key sets the printer format in the oximeter. When a new
oximeter is installed in the Docking Station, it may have a different print format. Always check the
printer format prior to pushing the start key.
Change modes by pushing the select key (
) until the desired mode is lit.
To start printing, push the print (
F
) key. The Print Mode Indicator on the printer will blink green
when printing is pending or in progress.
nOTe: The oximeter must be installed in the Docking Station to print.
nOTe: Thermal paper must be installed.
Mode
1
- Real Time Mode prints data every 5 seconds. See figure 7-7.
Mode
2
- Numeric Trend Data prints in a tabular form. See figure 7-8.
1
2
3
4
2008-JUN-22
REALTIME DATA
ID: 002_____________________
hh:mm:ss SpO2 Rate PI
(%) (bpm) (%)
18:38:15 97% 77 bpm 4.9%
18:38:20 97% 76 bpm 5.6%
18:38:25 97% 77 bpm 5.4%
18:38:30 95% 72 bpm 5.1%
18:38:35 94% 70 bpm 4.9%
18:38:40 95% 69 bpm 4.0%
18:38:45 96% 71 bpm 3.0%
18:38:50 96% 72 bpm 2.9%
18:38:55 96% 73 bpm 2.9%
Date when the printout
was started
Title
Patient Record number
Real time %SpO
2
, pulse
rate and PI values printed
every 5 seconds
1
2
3
4
1
2
3
4
5
2008-JUN-22
TREND DATA
ID: 009__________________________
Trend Start:2008-JUN-07 23:16:13
Trend End: 2008-JUN-07 23:16:48
hh:mm:ss SpO2 Rate PI
(%) (bpm) (%)
23:16:48 96% 93 bpm 7.5%
23:16:43 96% 94 bpm 8.3%
23:16:38 96% 90 bpm 8.8%
23:16:33 96% 89 bpm 9.2%
23:16:28 96% 89 bpm 9.6%
23:16:23 96% 87 bpm 8.7%
23:16:18 96% 87 bpm 6.0%
23:16:13 96% 90 bpm 2.9%
Date when the printout
was started
Title
Patient Record number
Start and end time for
trended data
Trended %SpO
2
, pulse
rate and PI values printed
every 5 seconds
5
1
2
3
4
Figure 7-7: Sample Real Time Data Log
Figure 7-8: Sample Numeric Trend Data
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......