Chapter 5: Operating Instructions
Home Use Operation Manual
5-1
Chapter 5: Operating Instructions
Powering the Oximeter
The WW1030 oximeter will operate from battery power or from external power with battery
back up. The optional WW1095 (30 Watt) AC power supply may be used to provide power to the
oximeter. The AC power supply is required when utilizing the docking station to ensure the proper
operation of the docking station with all accessories, including auxiliary battery charger and
optional printer.
WARnInG! The WW1030 will not operate without batteries installed. Properly charged
batteries provide a reserve source of power in case of external power failure.
never use an oximeter with discharged batteries to monitor a patient, as the
monitor may not operate properly or may fail to alarm in the case of external
power failure.
The WW1030 oximeter can obtain external power in the following ways:
The oximeter can be placed in its docking station. See
•
Chapter 7: Optional Docking Station
and Printer
.
The AC power supply can be plugged directly into the oximeter. See Figure 5-3.
•
The AC power supply can be plugged into the WW1089 USB Interface cable (Figure 5-4):
•
The oximeter can be powered by a PC through the USB Interface cable (Figure 5-5).
•
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......