S E C T I O N 1 2
•
M a i n t e n a n c e
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
43
peroxide solution and 1,260 ml of sterile water). The reservoir
holds 1.4 liters.
Maintenance Performed Every 12 Months
Disinfect the Reservoir
1
Place a container under the drain valve of the Fluid Warmer.
2
Drain the recirculating solution by turning the drain valve
clockwise 90 degrees.
3
When all solution has drained from the reservoir, close the drain
valve.
4
Remove the fill-port plug on the reservoir.
5
Prepare a 0.3% hydrogen peroxide/sterile water solution.
Mix 140 ml of 3% hydrogen peroxide solution and 1,260 ml of
sterile water.
6
Fill the reservoir with 1.4 liters of 0.3% hydrogen peroxide/sterile
water solution.
7
Replace the fill-port plug.
8
Insert a Disposable Set into the Fluid Warmer.
9
Turn the Fluid Warmer ON. Let the solution circulate for a
30-minute disinfection period.
10
Turn the Fluid Warmer OFF.
11
Empty the reservoir; c
lose the drain valve.
12
Remove the Disposable Set and discard according to
established hospital procedures.
13
Refill the reservoir to its
1.4-liter
capacity.
• If using distilled water, refill the reservoir with distilled water.
• If using a peroxide solution, refill the reservoir with a 0.3%
hydrogen peroxide/sterile water solution.