S E C T I O N 1 0
•
T r o u b l e s h o o t i n g
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer| Operator’s Manual
35
SECTION 10
Troubleshooting
Only qualified personnel authorized by the institution should perform any
routine maintenance and repairs to the Level 1
®
NORMOFLO
®
H-1129
Irrigation Fluid Warmer.
The following two tables feature general troubleshooting information
along with slow flow rate troubleshooting.
General Troubleshooting Guide
Problem
Check the following
No Power
Check to see if unit is plugged in and power is turned ON.
Be sure the unit is plugged into a working MAINS receptacle.
Disposables
Alarm activates
Check to see that the Heat Exchanger and Disposable Set
are properly installed.
Add
Recirculating
Solution Alarm
activates
Fill reservoir to the maximum reservoir level with recirculating
solution.
Over
Temperature
Alarm activates
Press OFF button to clear alarm. Then, press ON button to
power on. If the Fluid Warmer continues to alarm,
discontinue use of the medical device and remove from
service. Contact Smiths Medical or an authorized
representative for service.
Hot Cabinet
Check the air inlet on the bottom of the unit. Remove any
blockage or dust to insure adequate air flow.
Heat Exchanger
hard to install
Lubricate O-Rings in #1 and #2 block heat exchanger
sockets with silicone lubricant. Silicone lubricant part
# 80-04-002.
Loud
Compressor
Verify pneumatic tubing is fully seated into fittings.
Green Automatic
Operating LED
doesn't light up
during setup
Verify unit is plugged into MAINS.
If still no LED illumination, discontinue use of the medical
device and remove from service.