Mode Characteristics
• The Fluid Warmer must be ON for the Pressure Chamber to
work.
• Pressure is applied to the fluid bag in the Pressure Chamber
according to the amount specified by the Pressure Regulator
Knob (
d
).
• Pressure is indicated on the Pressure Gauge (
a
).
To exit this mode
• Push the Pressure Chamber Toggle Switch (
b
) to the minus (
-
)
position.
Unpressurized Mode
• Disable pressurization by turning the Pressure Chamber Toggle
Switch (
b
) on the Pressure Chamber to the minus (
-
)
unpressurized position.
Mode Characteristics
• Pressure is released from the Pressure Chamber.
• A fluid bag can be removed or loaded into the Pressure Chamber.
Pressure Regulation Mode
Pressure Regulator Knob
• The Pressure Regulator Knob (
d
) functions when the device is
ON and the Pressure Chamber Toggle Switch (
b
) is in the plus
(+) position. The knob lets you adjust the pressure in the
Pressure Chamber.
• The Pressure Regulator Knob controls both Pressure Chambers if
both are activated.
Pressure Dial Gauge
• When the Toggle Switch is in the plus (+) pressurized position
and the Fluid Warmer is ON, this gauge (
c
) displays the selected
pressure as you adjust the Pressure Regulator Knob.
• The Pressure Dial Gauge controls both Pressure Chambers if both
are activated.