S E C T I O N 7
•
C o n t r o l s a n d D i s p l a y s
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
19
Pressure Gauge Box
1
Pressure Dial Gauge
When pressure is activated, the Pressure Dial Gauge displays the
selected pressure as you adjust the Pressure Regulator Knob (
b
).
2
Pressure Regulator Knob
When pressure is activated, the Pressure Regulator Knob adjusts
the level of pressurization. As you turn the knob, the Pressure
Dial Gauge displays the selected pressure.
Interlocks
The Fluid Warmer has three Interlocks that detect for proper
installation of a Disposable Set's components.
Refer to the figure to identify the positions of Interlocks.
Note: Block
1
is not an Interlock. It cannot detect if a Disposable Set is
not correctly installed. It is identified here because it is an essential step
for proper installation of the Disposable Set components.
The three Interlocks are located on the Fluid Warmer and check for
proper installation of:
•
2
Heat Exchanger (top end)
•
3
Gas Vent
•
4
Heat Exchanger (guide)
Interlocks
2
,
3
, and
4
prevent the Fluid Warmer's pump from
circulating reservoir solution if the Disposable Set's Heat Exchanger and
Gas Vent are not installed properly.
If Check Disposables Alarm is Activated on the Fluid Warmer
a
Check Heat Exchanger for proper installation in Block
1
, and
Interlocks
2
and
4
.
b
Press Heat Exchanger down firmly in Block
1
to secure in
O-Ring.
c
Press Interlock
2
tab down firmly to engage the Interlock switch.
d
Check Gas Vent installation in Interlock
3
.
3
3