RENASYS TOUCH 

 Quick Reference

 

(refer to the User Manual for full details on RENASYS TOUCH)

- To switch between Continuous and Intermittent therapy, ensure therapy is paused and press the therapy mode toggle on the home screen

Intermittent therapy home screen

Continuous therapy home screen

  Step 1:

 Ensure that the device is powered on and 

the Continuous therapy home screen is displayed

  Step 2:

 Use the + or – buttons to select prescribed 

pressure setting

  Step 3: 

Press the Start/Pause therapy button to 

start delivering therapy

  When therapy is active the status indicator on top 

of the device illuminates green and the therapy 

indicator at the top of the screen will rotate orange

  Step 1: 

Ensure that the device is powered on and 

the Intermittent therapy home screen is displayed

  Step 2:

 Change the High and Low pressure 

settings and cycle times by selecting the setting 

and use the + or – to the prescribed pressure/

time

  Step 3:

 Press the Start/Pause therapy button to 

start delivering therapy

 

  When therapy is active the status indicator on top 

of the device illuminates green and the therapy 

indicator at the top of the screen will rotate orange

  All 4 settings can be adjusted within the following pressure ranges:
 

High Therapy 

  25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg

  High Cycle times:

 3, 5, 8, 10 minutes

 

Low Therapy 

  0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg

  Low Cycle times:

 2, 3, 5, 8, 10 minutes

3. Starting therapy (Intermittent)         

2. Starting therapy (Continuous)

1. Switching the device between Continuous and Intermittent therapy

Therapy mode toggle

Therapy mode toggle

Therapy set point

Status indicator

Status indicator

Increase value

Start/Pause therapy

Decrease value