RENASYS TOUCH
–
Quick Reference
(refer to the User Manual for full details on RENASYS TOUCH)
- To switch between Continuous and Intermittent therapy, ensure therapy is paused and press the therapy mode toggle on the home screen
Intermittent therapy home screen
Continuous therapy home screen
Step 1:
Ensure that the device is powered on and
the Continuous therapy home screen is displayed
Step 2:
Use the + or – buttons to select prescribed
pressure setting
Step 3:
Press the Start/Pause therapy button to
start delivering therapy
When therapy is active the status indicator on top
of the device illuminates green and the therapy
indicator at the top of the screen will rotate orange
Step 1:
Ensure that the device is powered on and
the Intermittent therapy home screen is displayed
Step 2:
Change the High and Low pressure
settings and cycle times by selecting the setting
and use the + or – to the prescribed pressure/
time
Step 3:
Press the Start/Pause therapy button to
start delivering therapy
When therapy is active the status indicator on top
of the device illuminates green and the therapy
indicator at the top of the screen will rotate orange
All 4 settings can be adjusted within the following pressure ranges:
High Therapy
25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
High Cycle times:
3, 5, 8, 10 minutes
Low Therapy
0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg
Low Cycle times:
2, 3, 5, 8, 10 minutes
3. Starting therapy (Intermittent)
2. Starting therapy (Continuous)
1. Switching the device between Continuous and Intermittent therapy
Therapy mode toggle
Therapy mode toggle
Therapy set point
Status indicator
Status indicator
Increase value
Start/Pause therapy
Decrease value