S E C T I O N 8
•
T r o u b l e s h o o t i n g
HOTLINE
®
3 Blood and Fluid Warmer | Operator’s Manual
23
Hot cabinet
Check for blocked air vents on the bottom or the back of
the HOTLINE
3 Warmer.
Note: Room temperature above 38
°
C may cause the
HOTLINE
3 Warmer to shut down and the Over
Temperature alarm to activate. In this situation, turn
the power switch OFF and allow the HOTLINE
3
Warmer to cool down before returning it to service.
Recirculating
solution leaks at the
socket where the
HOTLINE
3 Fluid
Warming Set plugs
into the HOTLINE
3
Warmer
Replace the HOTLINE
3 Fluid Warming Set.
Electrical
interference -
receiving or
transmitting
1 Move the HOTLINE
3 Warmer away from the device
in question.
2 Plug the HOTLINE
3 Warmer into a separate
electrical circuit.
• If the problem continues, notify Smiths Medical or
your local Smiths Medical distributor.