-91-
No.PS
※※
-OMX0002
○
Error indication function
This function is to display error location and content when a problem or error has occurred.
Error
Error displayed
Description
Measures
Over current
error
The load current applied to the switch
output has exceeded the maximum
value.
Turn the power off and remove the
cause of the over current.
Then supply the power again.
Residual
pressure error
During zero clear operation, pressure
greater than ±7% F.S. is present. Note
that the mode is returned to
measurement mode automatically
1 second later. The zero clear range
varies by ±1% F.S. due to variation
between individual products.
Release the applied pressure to
atmospheric pressure, and retry the
zero clear operation.
Pressurizing
error
Pressure exceeding the upper limit of
the set pressure range is applied.
Reset applied pressure to a level
within the set pressure range.
Pressure exceeding the lower limit of
the set pressure range is applied.
System error
Displayed if an internal data error has
occurred.
Turn the power off and on again.
If the failure cannot be solved,
contact SMC.
Version does not
match
Version of master and IO-Link
does not match.
Mismatch because master version is
1.0.
Align the master IO-Link version to
the device.
If the error cannot be reset after the above measures are taken, or errors other than above are displayed,
please contact SMC.
Содержание ISE20CH-L
Страница 96: ...95 No PS OMX0002 Dimensions 02 type...
Страница 97: ...96 No PS OMX0002 F02 type C01 type A2 type B2 type...
Страница 98: ...97 No PS OMX0002 02L type...
Страница 99: ...98 No PS OMX0002 F02L type C01L type A2L type B2L type...
Страница 100: ...99 No PS OMX0002 Bracket mounting dimensions Bracket A...
Страница 101: ...100 No PS OMX0002 Bracket C...
Страница 103: ...102 No PS OMX0002 Mounting dimensions of panel mount adapter bottom piping type...
Страница 104: ...103 No PS OMX0002 Mounting dimension of panel mount adapter bottom piping type Front protective cover...
Страница 106: ...105 No PS OMX0002 Vertical...