Chapter 3 Transport and Setting Up
3.3
Installation
HRSE Series
3-8
3.3.3
Preparation and wiring of power supply cable
Preparation
【
For power supply : 200/230V
】
1.
Strip the sheath from the cable of the product.
2.
Connect the other end of the cable to a terminal (e.g. crimped terminal) that is compatible to the
secondary side of the earth leakage breaker of the user's equipment.
Fig. 3.3-3 Power supply cable
Cable
The electrical facilities should be installed and wired in
accordance with local laws and regulations of each country and by
a person who has knowledge and experience.
Check the power supply. Operation with voltages, capacities and
frequencies other than the specified values can cause fire and
electrical shock.
Wire with an applicable cable size and terminal. Forcibly mounting
with an unsuitable size cable may result in heat generation or fire
Crimped terminal
(Prepared by customer)
L (White)
Earth leakage
breaker
(Secondary side)
N (Black)
E (Green)