Tracer

VM

Operating Instructions 

Applies to VM Flowmeter models. 

General

The Tracer

VM

Electronic Flowmeter is a non-display flow and

temperature sensor that provides a 0.5 - 3.5 Volt output only. It is
flow-direction specific. Refer to the arrow on the body for correct
flow direction. Flow in the opposite direction of the arrow will yield
inaccurate voltage output. Flow sensor technology is vortex shedding
behind a bluff body.

Temperature

Operating Range ..........................................0 to 100°C (32 to 212°F)
Temperature Equation.......Temperature (°C) = [(100/3) * V] - (50/3)

Temperature (°F) = (60 * V) + 2

Maximum Pressure...................................................10.3 bar (150psi)

Power

Power Supply Requirements ............................5VDC ±5% (external)

max. 10 mV ripple, 50 Hz

min. output current 10mA

separated from hazardous live circuitry by

double or reinforced insulation

power limitation: 150VA; current limitation: 8A

Output signals: ................................................................. Ratiometric
Flow Signal  .............................................. 0.5 - 3.5V (zero at 0.35V)
Temperature Signal............................................................. 0.5 - 3.5V
Power Consumption.............................................................. < 50mW
Load Impedance ..................................................................... > 10kΩ

Specifications

Flow

Accuracy ....................±5% of Full Range

Flow Ranges and Equations (V = Voltage)

1-15 LPM

(LPM) = [(14/3) * V] - (4/3)

.3 to 4 GPM

(GPM) = [1.233 * V] - 0.352

2-40 LPM

(LPM) = [(38/3) * V] - (13/3)

.5 to 10.6 GPM

(GPM) = [3.347 * V] - 1.145

5-100 LPM

(LPM) = [(95/3) * V] - 10.833

1.3 to 26.4 GPM

(GPM) = [8.365 * V] - 2.862

10-200 LPM

(LPM) = [(190/3) * V] - (65/3)

2.6 to 52.8 GPM

(GPM) = [16.731 * V] - 5.724

0.35V output for all ranges is equal to zero flow

Component Materials

Sensing Element.... Silicone-Based MEMS Sensor
Seal (sensor to housing) ................. EPDM Rubber
Insert..................................................... PPA 40 GF
3/8" & 1/2" Body Size..............Glass-Filled Nylon 

Flow Body with

Brass or Nylon End Caps

3/4" thru 1-1/2" Body Size ....Anodized Aluminum

or Stainless Steel Flow Body

Cable ....................................4-Conductor, 24AWG

2.9M (9.5ft)

Directives

Flow sensors are in conformity with these Council
directives on the approximation of the laws of the
EC member states:

• Low Voltage Directive (2006/95/EC)

- Standards used: EN 61010-1:2001

• EMC Directive (2004/108/EC)

- Standards used: EN 61326-1:2006 and 61326-
2-3:2006

Smartflow Vortex flow sensors fall under Article 3,
§3 of PED Directive 97/23/EEC and are therefore
not required to be CE-marked according to this
directive.