background image

TL-2016REV Digital Wireless Transmitter

Quick Start

 Instructions

Need Set-up Help?

Call Our Friendly Customer Service Reps at (800) 650-3637

TL-2016REV Digital Wireless Transmitter with Voice & Tone Options

1229 N. McDowell Blvd., Petaluma, CA. 94954

[email protected]

www.smartcaregivercorp.com

Toll Free: (800) 650-3637

Step 4

Warning:

 Failure to comply with warning may result in injury or death. This device is not suit-

able for all individuals. Other devices may be required. This device is not a substitute for visual 

monitoring by caregiver. The manufacturer does not claim that this device will stop elopement 

and or stop falls. This device is to help augment caregivers comprehensive resident mobility 

management program. Test this device before each use. Read the instructions and Legal Dis-

claimer. Please be aware that the transmitter will not be 100% accurate if it is out of range at anytime. 

It is the end-user’s responsibility to make sure that the product is used correctly and within range of 

each other.
Seller warrants that the goods sold will be free from defects of workmanship and manufacturing, for 

a period of one year from the date of sale. Other than this limited warranty, seller makes no express or 

implied warranties. Except for this limited warranty, NO EXPRESS OR IMPLIED WARRANTIES ARE GIVEN 

AND NO AFFIRMATION OF SELLER, BY WORDS OR ACTION, WILL CONSTITUTE A WARRANTY. THE 

GOODS BEING PURCHASED ARE BEING SOLD ON A “WITH ALL FAULTS” BASIS. THE ENTIRE RISK AS TO 

THE QUALITY AND PERFORMANCE OF THE GOODS IS WITH THE BUYER. SHOULD THE GOODS PROVE 

DEFECTIVE FOLLOWING THEIR PURCHASE, THE BUYER, AND NOT THE MANUFACTURER, DISTRIBUTOR 

OR RETAILER, ASSUMES THE ENTIRE COST OF ALL NECESSARY INSTALLATION, SERVICING OR REPAIRS.
The Smart Caregiver Corporation Digital Wireless Transmitter is designed to be installed by the end-

user. As such, it is the entire responsibility of the buyer to insure that the system is properly installed 

and tested. Further, the system is not designed to replace good care giving practices including, but 

not limited to:
a.) Direct patient supervision
b.) Adequate training for staff personnel for elopement
c.) Testing of the system before each use.
WHERE PERMISSIBLE, IT IS THE SELLER’S INTENT TO LIMIT ANY LIABILITY FOR SPECIAL, CONSEQUEN-

TIAL OR INCIDENTAL ECONOMIC DAMAGES TO REFUND OF THE PURCHASE PRICE OR REPLACEMENT 

OF THE GOODS. SELLER DISCLAIMS ANY LIABILITY FOR COMPENSATORY OR OTHER DAMAGES ARIS-

ING OUT OF THE USE OF THE GOODS.

WARNING: The manufacturer does not claim that this device will 

stop elopement and or stop falls. This device is to help augment 

caregivers comprehensive resident mobility management 

program. This product is to be used by a licensed caregiver only.

Step 1

Step 2

Step 3

The switches above can be found inside the 

battery cover at the back of the unit. You will need 

a small screwdriver to adjust transmitter settings.

• Insert 9v battery (not included) into battery compartment. (See first 
diagram).

• Set delay switch (0 or 1 second). A one second delay can usually 
allow for normal movement of resident when using a pressure pad. 
(See second diagram).

• Set tone switch. Choose from a single chime (1), a siren alert (2), or 
“Westminster” chimes (3). (See second diagram).

• Set volume. Choose from “high”, “medium”, or “silent”. Use “silent” if 
you want alarm to signal at a Central Monitor Unit or Wireless Receiver 
and not in the room. (See second diagram). Please note that the 
monitor will be shipped with the alarm volume stet at 90 decibels. (90 
decibels is the legally accepted occupational noise allowance).

• Verify code programming switch. The TL-2016REV is sent out 
pre-programmed for use with the TL-2017REV. If you wish to use 
TL-2016REV with a central monitoring unit or change programming 
in any way, follow the programming instructions on the back page. 
Otherwise, user may skip to Step 3.

• Select voice response (optional) by moving dial to position 0, 1, or 
2. Choose from voice settings of “Sit back down.” (0); “Lay back down.” 
(1); or warning tone (2).

• Connect external device (i.e. pressure pad for bed or chair, floor 
mat, etc). Lace cord through T-shaped strain relief to guard against 
accidentally unplugging device or general cord damage. (See third 
diagram). Make sure the cord is placed so it is well out of resident’s 
reach and is not a tripping hazard.

• Test system by triggering external device (i.e. pressure pad, floor mat, 
etc). TL-2016REV should alarm (unless you have set it to “silent” as in 
step four). Receiver device (TL-2017REV or Central Monitoring Unit) 
should alarm. Test system before every use.

Step 5

• Mount TL-2016REV out of resident’s reach if you are using an external 
device. Mount TL-2016REV within resident’s reach if you intend to use 
the “nurse call” feature. The TL-2016REV also has a 1/4” input for a nurse 
call button extension cord. 

• A setting of “0” can be received as a common address by all of any number 

of central monitors. During programming, if no more activity is detected 

for approximately 1 minute, the unit will automatically return to normal 

operating mode and any incomplete programming must be adjusted again, 

from the start. If a central monitor is included, it should be programmed 

to receive the address code from designated monitors. A setting of “0” at a 

central monitor (if included) will receive calls from any bedside monitor unit 

within wireless range.

• When using a central monitor, the unit must be programmed to the bed 

number, also to the room number. The area code ( in step 8) must be set to 

match the programming for the associated wireless central monitor.

• If using the wireless remote chime and/or the wireless indicator light, the 

room number must be matched with the programming for the associated 

wireless remote chime and light. 

Note

+

-

Отзывы: