(Model J) Page 15
2.2 Pneumatic System
The pneumatic system comprises of the tubing and electro-mechanical valves necessary to
provide the gas in conventional and oscillatory ventilation modes. The two gas controlling
functions are blending and pressure generation
2.2.1 Blending
The method used for blending air and oxygen in known proportions is to pressure regulate
the two supplies (air and oxygen) so they produce equal flow rates and then allow each
supply into a mixing chamber for a time period equivalent to the proportions required. For
example, delivering oxygen at a set flow rate into a mixing chamber for 1 second and air at
the same flow rate for 2 seconds will result in a mixture of 1 part oxygen to 2 parts air
(resulting in a mix of 47.3%).
2.2.2 Pressure Generation
There are three nozzles within the exhalation block in the pneumatic subsystem. One for
generating negative pressure in the patient circuit and the other two for generating positive
pressure. The pressures generated from the three nozzles are controlled by three
electronically controlled pressure regulators. The negative and one of the positive nozzle
pressures can also be switched on and off rapidly with in-line (high speed) solenoid valves.
The other positive nozzle (the mean jet) is used to generate steady pressures in ventilation
(CPAP or PEEP pressures in conventional modes and mean pressures in HFO modes).
These three nozzles (or jets) are used in various combinations to generate all ventilation
modes.
2.2.3 Conventional ventilation
In non HFO modes the negative (or reverse) jet is used in a steady mode to provide a small
amount of flow to offset the inadvertent patient circuit pressure generated from the fresh gas
flow of 8 lpm. The mean jet is also used in a steady mode to generate the base pressure
level (CPAP or PEEP) and the forward jet is used to generate the PIP pressure during the
inspiratory phase. The rise time of the inspiratory phase is controlled by dynamically
controlling the forward jet pressure regulator rather than switching a steady pressure with
the high speed valves.
This provides a smooth rise in pressure and allows user adjustable rise times rather than
abrupt changes and pressure “ringing” which can result from high speed switching. The fall
of the inspiratory wave is also controlled by the forward jet pressure regulator to bring the
pressure down quickly and smoothly, using the high speed valves to do this would again
result in inadvertent ‘ringing’ which causes difficulties for the monitor subsystem in trying to
detect a patient breath attempt by monitoring the pressure alone. Once the pressure has
been brought close to the base pressure, after about 100 ms, the forward jet solenoid is
switched off to prevent any further artefact causing false triggering.
Содержание SLE5000 Model J
Страница 1: ...SLE5000 Model J service manual Version 4 3 1 4 3 2 software ...
Страница 8: ...Page 8 Model J This page is intentionally blank ...
Страница 9: ... Model J Page 9 Warnings and Cautions ...
Страница 13: ... Model J Page 13 Principles of Operation ...
Страница 17: ... Model J Page 17 Description of Symbols ...
Страница 20: ...Page 20 Model J This page is intentionally blank ...
Страница 21: ... Model J Page 21 Equipment List ...
Страница 23: ... Model J Page 23 Engineering Mode Software ...
Страница 31: ... Model J Page 31 Component Replacement ...
Страница 54: ...Page 54 Model J This page is intentionally blank ...
Страница 55: ... Model J Page 55 A0763 02 M Board ...
Страница 58: ...Page 58 Model J This page is intentionally blank ...
Страница 59: ... Model J Page 59 Electrical Safety Testing ...
Страница 60: ...Page 60 Model J This page is intentionally blank ...
Страница 62: ...Page 62 Model J This page is intentionally blank ...
Страница 63: ... Model J Page 63 PSU Testing ...
Страница 64: ...Page 64 Model J This page is intentionally blank ...
Страница 66: ...Page 66 Model J This page is intentionally blank ...
Страница 67: ... Model J Page 67 Power Up Down Battery Care ...
Страница 72: ...Page 72 Model J This page is intentionally blank ...
Страница 73: ... Model J Page 73 Maintenance and Overhaul ...
Страница 75: ... Model J Page 75 12 36 Month Maintenance ...
Страница 82: ...Page 82 Model J This page is intentionally blank ...
Страница 83: ... Model J Page 83 24 48 Month Overhaul ...
Страница 85: ... Model J Page 85 24 Month 10 000 Hour Overhaul ...
Страница 92: ...Page 92 Model J This page is intentionally blank ...
Страница 93: ... Model J Page 93 48 Month 20 000 Hour Overhaul ...
Страница 107: ... Model J Page 107 Calibration Procedure V4 3 1 V4 3 2 ...
Страница 128: ...Page 128 Model J This page is intentionally blank ...
Страница 129: ... Model J Page 129 Functional Testing ...
Страница 161: ... Model J Page 161 Trouble shooting ...
Страница 170: ...Page 170 Model J This page is intentionally blank ...
Страница 171: ... Model J Page 171 Software Version History ...
Страница 173: ... Model J Page 173 Technical Information ...
Страница 218: ...Page 218 Model J 26 Ventilator labelling ...
Страница 219: ... Model J Page 219 Schematic Circuit Diagrams ...
Страница 220: ...Page 220 Model J 27 Electrical Block Diagram ...
Страница 222: ...Page 222 Model J 28 2 Schematic ...
Страница 223: ... Model J Page 223 29 Main Loom Circuit Diagram ...
Страница 224: ...Page 224 Model J 30 L0275 OEM2 Assembly Schematic ...
Страница 225: ... Model J Page 225 A0761 Circuit Diagram ...
Страница 226: ...Page 226 Model J 31 A0761 circuit diagram 31 1 Transducer PCB Assembly ...
Страница 227: ... Model J Page 227 A0763 02 Circuit Diagrams ...
Страница 228: ...Page 228 Model J 32 A0763 02 circuit diagrams 32 1 Micro Controller ...
Страница 229: ... Model J Page 229 32 2 Conventional Valve Drive ...
Страница 230: ...Page 230 Model J 32 3 Power Distribution And Hardware Identifier ...
Страница 231: ... Model J Page 231 32 4 Analogue Valve Drive Non Volatile Memory Interboard Comms ...
Страница 232: ...Page 232 Model J 32 5 High Speed Valve Drive ...
Страница 233: ... Model J Page 233 32 6 Analogue Data acquisition Monitor port ...
Страница 234: ...Page 234 Model J 32 7 Control Monitor PCB PSU Circuit ...
Страница 235: ... Model J Page 235 32 8 Control Monitor Micro Non isolated ...
Страница 236: ...Page 236 Model J 32 9 Control Monitor Micro Isolated ...
Страница 237: ... Model J Page 237 32 10 Control Monitor Analog Isolated ...
Страница 238: ...Page 238 Model J 32 11 Control Monitor Alarm Controller ...
Страница 239: ... Model J Page 239 Illustrated Parts List ...
Страница 242: ...Page 242 Model J Ventilator assembly Figure 1 Sheet 1 of 5 A 5 10 1 View A B ...
Страница 244: ...Page 244 Model J Ventilator assembly Figure 1 Sheet 3 of 5 115 110 115 120 125 ...
Страница 245: ... Model J Page 245 Ventilator assembly Figure 1 Sheet 4 of 5 145 160 120 135 150 130 140 155 ...
Страница 246: ...Page 246 Model J Ventilator assembly Figure 1 Sheet 5 of 5 170 175 165 180 ...
Страница 249: ... Model J Page 249 Base plate and side door assembly Figure 2 30 20 5 15 10 35 40 25 ...
Страница 251: ... Model J Page 251 Front facia assembly Figure 3 10 5 70 45 50 20 15 40 30 35 65 75 25 ...
Страница 253: ... Model J Page 253 Electronic module Figure 4 Sheet 1 of 13 1 A ...
Страница 254: ...Page 254 Model J Electronic module Figure 4 Sheet 2 of 13 6 10 15 5 25 30 View A View B 20 10A ...
Страница 255: ... Model J Page 255 Electronic module Figure 4 Sheet 3 of 13 45 35 C 40 ...
Страница 260: ...Page 260 Model J Electronic module Figure 4 Sheet 8 of 13 View G 445 495 500 490 ...
Страница 262: ...Page 262 Model J Electronic module Figure 4 Sheet 10 of 13 View J 565 565 570 575 K L 595 590 605 610 600 580 580 ...
Страница 273: ... Model J Page 273 Pneumatic module Figure 5 Sheet 1 of 7 1 B 5 10 15 20 25 35 30 40 65 45 A View A 50 60 55 85 80 75 70 ...
Страница 274: ...Page 274 Model J Pneumatic module Figure 5 Sheet 2 of 7 100 95 C View B 115 125 110 90 105 145 140 135 130 120 ...
Страница 276: ...Page 276 Model J Pneumatic module Figure 5 Sheet 4 of 7 View D 210 215 195 190 205 195 220 230 225 180 185 ...
Страница 277: ... Model J Page 277 Pneumatic module Figure 5 Sheet 5 of 7 265 255 245 235 260 250 250 F E 240 270 ...
Страница 285: ... Model J Page 285 DISS air and oxygen fittings Figure 6 Sheet 1 of 1 10 15 20 25 30 35 45 5 40 50 70 65 60 55 75 80 ...
Страница 287: ... Model J Page 287 Modifications ...
Страница 303: ... Model J Page 303 Appendix 1 Calibration Checklist ...
Страница 304: ...Page 304 Model J This page is intentionally blank ...
Страница 311: ... Model J Page 311 Appendix 2 Check list ...
Страница 312: ...Page 312 Model J This page is intentionally blank ...
Страница 317: ... Model J Page 317 Appendix 3 List of Technical Bulletins Service Information Letters ...
Страница 321: ... Model J Page 321 Appendix 4 RS232 ...
Страница 330: ...Page 330 Model J This page is intentionally blank ...
Страница 331: ... Model J Page 331 Appendix 5 VueLink Technical Data ...
Страница 344: ......