112
The following calculation formulae are available for EQ and EP:
Formula for linear and
dynamic titration to EQ1
Formula for titrations to end-
point (EP 1 and EP2)
Additional information
No formula
No result will be determined..
(EQ1-B)*T*M*F1/(W*F2)
(EP1-B)*T*M*F1/(W*F2)
Formula for calculating the
concentration of a sample taking into
account a blank value in terms of ml.
Direct titration to one EQ or EP1 (ex.::
chloride, p or m value)
(B
–EQ1)*T*M*F1/(W*F2)
(B-EP1)*T*M*F1/(W*F2)
Formula for calculating the
concentration of a sample taking into
account a blank value in terms of ml.
Reverse titration (examples. CSB,
saponification number)
(B*F3
–EQ1*F1)*T*M/(W*F2)
(B*F3
–EP1*F1)*T*M/(W*F2)
Formula for calculating the
concentration of a sample taking into
account a blank value, including a
multiplicative factor. Back titration.
(W*F2)/(EQ1-B)*M*F1)
(W*F2)/(EP1-B)*M*F1)
Formula for calculating a titer (T) of a
titration solution.
(W*F2)/(EQ1-B)*M*T*F1)
(W*F2)/(EP1-B)*M*T*F1)
Formula for calculating the
concentration of a sample taking into
account a blank value in ml. Direct
titration to one EQ or EP1.
(W*F2)/(B-EQ1)*M*T*F1)
(W*F2)/(B-EP1)*M*T*F1)
Formula for calculating the
concentration of a sample taking into
account a blank value in ml. Back
titration (NCO-value, Epoxy-number).
EQ1
EP1
Calculation of the consumption in the
equivalence or end point.
EP2*T*M*F1/(W*F2)
Formula for the calculation of
concentration of a sample. Direct
titration to 2 EP. Here EP2 (p and m
value)
(EP2-EP1)*T*M*F1/(W*F2)
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here calculation of
the difference between EP2-EP1.
(F3*EP2-EP1)*T*M*F1/(W*F2)
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here: calculation of
the difference between EP2-EP1,
taking into account a multiplicative
factor for EP2.
(F1/W) * EP1 *F2
Calculation of the des TAC (
T
otal
A
norganic
C
arbonat reserve)
((F1/W)*(EP2-EP1) * F3-F4)*F5
Calculation of the FOS (
V
olatile
O
rganic
A
cids)
FOS/TAC-value
The abbreviations used here have the following meaning:
ml:
Total consumption, e.g. for pH Stat
EQ:
Consumption at the equivalence point 1 in ml
EP:
Consumption at the end point 1 and 2 in ml
B:
Blank value in ml. Mostly determined by way of titration
T:
Titer of the titration solution (e.g. 0.09986)
M:
Mol; mol- or equivalence weight of the sample (e.g. NaCl 58.44)
F1-F5
Factor 1-5. conversion factors
W
“Weight“, weighed-in quantity in g or volume in ml.”