4 Calibration
4.1 Calibration of poultry scale
This section is relevant only to DOL 53X with software version 6.6 or higher.
The scale should be calibrated at least once a year.
When calibrating equipment, it must have a fixed, stable temperature. Therefore, if you move the equipment into
the livestock house, it requires some time to become acclimatized.
If you connect the scale to a house controller with software version 6.6 or later, it should only be calibrated on
the controller - not on the scale itself.
Display reading
Functional explanation
Start calibration
Activating the calibration.
The house controller returns to normal mode
OFF
, once the calibration is com-
pleted.
Calibration state
Idle
Wait for stable empty scale
Place calibration weight
Remove calibration weight
Calibration finished
Reading of the calibration sequence.
Calibration weight
Adjusting the exact weight used for calibration.
Calibration date
Date of the last calibration.
Calibration deviation
Deviation from previous calibration in percentages.
Voltage
Displays the current input voltage from the scale.
Before calibration, it is important to ensure that the animals do not have access to the scale that needs calibra-
tion. For example, place a bucket over the scale.
In the menu
Technical | Calibration | Production | Bird
scale
.
Enter the used calibration weight. See table below for calibra-
tion weight requirements for the different types of weights.
Activate
Start calibration
to start the calibration.
Follow the status of the calibration through the following steps:
Wait for stable empty scale
Place calibration weight:
Place the calibration weight on the
scale and wait 30 sec.
Remove the calibration weight:
Remove the calibration weight
and wait 30 seconds while the scale is empty. This is to ensure
that the scale remains empty until the calibration is completed.
Calibration finished
.
Type of scales
Recommended calibration weight
[kg]
5, 10, 20 and 30 kg scale 5