Protection
IMx-M Manager
SKF Multilog On-Line System IMx-M
2 - 91
User Manual
Utilities
Through Utilities, it is possible to set the rack into arm mode (
ARM Rack
) while in
Maintenance mode to allow testing the complete measurement chain from sensor to
relay output. This will allow to test proper relay operation using simulated inputs, e.g.
simulate an Alert condition.
It is also possible to reset latching alarms (
Reset Latched
) and peak speed (
Reset Peak
Speed
) while in Maintenance mode.
Figure 2 - 65.
Maintenance View
–
Utilities.
Acceptance Test
The aim of Acceptance Test is to produce a report which shows that all channels and
relays were checked for correct values and functionality.
Figure 2 - 66.
Maintenance View - Acceptance Test.