Hardware
Installation
4 - 12
SKF Multilog On-Line System IMx-M
User Manual
Figure 4 - 7.
Power Fuses.
Power Fuses
The power fuses are slow blow 6,3 A (6,3 AT 250 V, 5 x 20 mm).
After operation of the fuse, due to mains power double pole (neutral) fusing, parts of
the equipment that remain energized might represent a hazard during servicing.
Important - Caution to a possible hazard due to mains power double
pole/neutral fusing.
Inrush Current
The maximum inrush current is 70 A.
Sensor Power Output
Each channel, both analog and digital, has built-in individual 24 V (23.5
28 V) sensor
power supply.
The maximum current output is 30 mA per channel.