The 231 and 237 Series sample bags feature dual stainless steel 

fi

 

tt

 ings – an inlet 

valve for 

fl

 ushing and 

fi

 lling and a septum 

fi

 

tt

 ing for removing samples.

Guidelines for Bag Sampling

1.  Ensure that the bag material and 

fi

 

tt

 ings are appropriate for the compounds 

to be sampled.

2.  Use only PTFE tubing to connect the sample bag to the pump to prevent sample 

loss by adsorption on tubing walls.

3. Before using, 

fl

 ush the bag thoroughly with puri

fi

 ed air or nitrogen.

4.  Analyze the sample within 24 to 48 hours. Long-term storage of air-contaminant 

mixtures in bags is not recommended.

5.  Do not ship sample bags by air unless the cargo cabin is pressurized.

Notes and Cautions

•  Maximum Operating Temperature — 

Do not exceed these speci

fi

 cations:

•  231 Series Bags:

 225 F (107 C) due to 

fi

 

tt

 ing O-ring temperature rating

•  237 Series Bags:

 140 F (60 C) due to the SamplePro

®

 FlexFilm bag material 

 temperature 

rating

•  Do not use sample bags to sample compounds with boiling points > 249.8 F (121 C).

•  SKC sample bags are designed for sampling air at atmospheric pressure only. A

tt

 empting 

to pressurize the bag can result in bag rupture and sample loss. Do not ship bag samples 
by air freight in a non-pressurized cargo cabin. Bags can burst under such conditions.

•  All federal and state packaging and transporting regulations apply.

• SKC bag 

fi

 

tt

 ing valves are extremely durable but are not intended for use as handles or 

hanging devices. This type of handling may damage the seal, causing leakage. This is 
considered to be misuse and will void the SKC warranty.

•  Failure to follow warnings and cautions voids any warranty.

•  Sample bags are designed for single use only.

Содержание 231 Series

Страница 1: ...rgo cabin is pressurized Notes and Cautions Maximum Operating Temperature Do not exceed these specifications 231 Series Bags 225 F 107 C due to fitting O ring temperature rating 237 Series Bags 140 F 60 C due to the SamplePro FlexFilm bag material temperature rating Do not use sample bags to sample compounds with boiling points 249 8 F 121 C SKCsamplebagsaredesignedforsamplingairatatmosphericpress...

Страница 2: ...le compounds that exceed 140 F 60 C temperature Do not use sample bags to sample compounds with boiling points 249 8 F 121 C 4 When sampling is complete turn off the pump and loosen turn counterclockwise the knurled thumbscrew a maximum of 2 revolutions 5 Close the shut off valve by pulling up on the valve stem Note Ensure the PTFE tubing is not pulled off while pulling up on the valve stem Tighten t...

Страница 3: ...ve closed Incorrect Inflation Correct Inflation SKC sample bags are available in Tedlar FlexFoil PLUS and Standard FlexFoil with fittings made of stainless steel polypropylene or PTFE in sizes ranging from 0 5 to 100 liters SKC also manufactures custom sample bags Contact SKC at 724 941 9701 or skcorder skcinc com For sample bag stability reports visit http www skcinc com catalog pdf instructions ...

Страница 4: ... feet Cat No 231 9 21 PTFE Tubing fits Vac U Chamber sample inlet port 1 4 inch ID 5 16 inch OD 10 feet Cat No 231 937 50 feet Cat No 231 924 www skcinc com SKC Limited Warranty and Return Policy SKC products are subject to the SKC Limited Warranty and Return Policy which provides SKC s sole liability and the buyer s exclusive remedy To view the com plete SKC Limited Warranty and Return Policy go ...

Отзывы: