17
Turn the
fl
ow adjust screw (Figure 3, #2) for the appropriate port
of the low
fl
ow holder until the desired
fl
ow rate is achieved.
Turn clockwise to decrease the
fl
ow.
Do not adjust the
fl
ow on the pump. Adjust the
fl
ow only by
using the
fl
ow adjust screw on the low
fl
ow holder.
Do not exceed 500 ml/min
fl
ow rate per tube.
When the desired
fl
ow is set on the
initial tube, place the pump in Hold
by pressing Flow and Battery Check.
Remove the calibrator from the tube
and connect to the exposed end of
the next sorbent tube. Press Flow and
Battery Check and repeat the
fl
ow
adjustment process until all tubes are
fl
ow calibrated. Changing the
fl
ow on
one tube will not affect the
fl
ow rate through the remaining tubes.
Do not exceed 500 ml/min
fl
ow rate per tube.
For tri and quad models,
fi
rst rotate each anti-tamper cover
(Figures 3 and 4) to expose the
fl
ow adjust screws, then
adjust the appropriate screw until the calibrator indicates the
desired
fl
ow.
When the
fl
ow rate is set for each tube, press Flow and Battery
Check to place the pump in Hold and disconnect the calibrator.
Replace the sampling media used for calibration with unexposed
media for sample collection. Use a protective tube cover to
prevent tube breakage.
If sampling with fewer tubes than number of ports, insert unopened
sorbent tubes in the empty ports to seal them.
2
(cont’d)
Screw
Screw
Figure 4
Cut-away of Tri/Quad Low
Flow Holder