2.5
Allocation of acquisition system to patient
During routine practice, the clear allocation of the acquisition system to
the patient to be examined is ensured. This also applies to the
allocation of the X-rays to the patient data stored by SIDEXIS XG /
SIDEXIS 4.
2.6
Storage battery
General guidelines and warnings
Hazards may be caused by lithium-ion storage batteries. Incorrect
charging or improper use may result in fire as well as personal injury
and damage to property.
If the WiFi interface with the installed storage battery becomes
unusually warm or deformed, switch the unit off immediately and consult
your specialist dealer.
No liability can be accepted for the storage battery for damage resulting
from misuse, incorrect charging or improper use.
Charging the storage battery
Only use the charger and power supply unit supplied to charge the WiFi
interface storage battery. Otherwise the storage battery may overheat
and ignite.
The charging temperature should be between 0°C (32°F) and 45°C
(113°F) for an optimum charge.
Ensure that no flammable materials are nearby during charging. If the
WiFi interface with the storage battery is warm to the touch, let it cool
down first of all before charging it.
The storage battery is only half charged upon delivery.
Storage battery lifetime
The battery should be recharged in the charger between patients with
regular use of the WiFi interface. This prolongs the battery lifetime.
Where the WiFi interface is not in use for longer than one week (on
account of a vacation for example), it should be removed from the
charging device. However, please note that it still uses a low amount of
power even when switched off on account of internal switching
operations, and a fully-charged storage battery will be dead in approx. 5
weeks. The WiFi interface must be recharged at this point at the latest
in order to prevent an exhaustive discharge of the storage battery which
shortens the battery lifetime.
If the storage battery has been charged for several hours and shows an
"orange" charge state after just one exposure (see "Determine unit
status" [→ 66]), the storage battery must be taken out of service and
disposed of.
Storage and transport
Store storage batteries at room temperature between 4°C (40°F) and
21°C (70°F). Do not expose the storage battery to direct sunlight for
long periods of time. Do not store the storage battery at locations where
the temperature may exceed 49°C (120°F).
For temporary storage and transportation of the storage battery the
temperature should not fall below -7°C (20°F) or exceed 66°C (150°F).
Temperatures of 66°C (150°F) and above may damage the storage
battery where stored for a longer period (more than 2 hours) and may
cause fire.