4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
4.5.14.3.5
Setting the direction of rotation
The direction of rotation can be changed only with the motor stopped.
Counterclockwise rotation is performed without torque limitation. The torque
setting keys are hidden when counterclockwise rotation is selected.
9
The
Simple/Advanced Endodontics program
is displayed on the
touchscreen.
¾
Touch the
CCW Rotation
key on the touchscreen.
or
¾
Press the left button of the foot control.
ª
For counterclockwise rotation: The
CCW Rotation
key is highlighted
orange and an orange CCW arrow appears.
For clockwise rotation: The
CCW Rotation
key is displayed gray and the
orange CCW arrow disappears.
4.5.14.3.6
Sub-screen functions
Opening the endodontics sub-screen
Opening the endodontics sub-screen
9
The
Simple Endodontics progra
m
is displayed on the touchscreen.
1.
Touch the
Sub-screen
key.
ª
The
Advanced Endodontics program
is displayed.
NOTE: Audible warning signal during CCW rotation
After you start the electric motor with the foot control, an audible warning
signal sounds 6 times if counterclockwise rotation is activated.
NOTE: Simple/Advanced Endodontics program display
After the start of a treatment, the
Simple Endodontics program
is shown by
default. After setting down the instruments and retrieving them again, the last
active program, i.e., the
Simple
or the
Advanced Endodontics
program is
displayed.