4 Operation
Operating Instructions TENEO
Dentist element
4.5.13.3
Setting the intensity
Selecting the intensity with the quick setting keys
9
The SIROTOM HF electrosurgical handpiece is removed from its holder.
9
The
Electrosurgery screen
is displayed on the touchscreen.
¾
Touch one of the quick setting keys in the bottom line key briefly (< 0.5 s).
ª
The quick setting key is highlighted orange. The selected intensity is
displayed in percent in the first line (Intens).
Setting intermediate intensity values with quick setting keys
9
The SIROTOM HF electrosurgical handpiece is removed from its holder.
9
The
Electrosurgery screen
is displayed on the touchscreen.
¾
Increasing the intensity: Press and hold a quick setting key whose
intensity value is greater than or equal to the value displayed in the first
line (> 0.5 s).
Reducing the intensity: Press and hold a quick setting key whose
intensity value is less than the value displayed in the first line (> 0.5 s).
ª
The selected intensity is displayed in percent in the first line (Intens).
The quick setting keys are shaded gray for intermediate values.
Selecting the intensity with function levels
9
The SIROTOM HF electrosurgical handpiece is removed from its holder.
9
The
Electrosurgery screen
is displayed on the touchscreen.
¾
Set the intensity using the – and + keys.
< 0.5s coarse adjustment, > 0.5s fine adjustment
ª
The selected intensity is displayed in percent in the first line (Intens).
4.5.13.4
Setting the modulation type
In modulation type 0 (Mod 0), the HF output voltage (U
H
F
) is constant. In
Mod 1 to Mod 4 the output voltage is increased further and pulsed. The
maximum output power remains limited to 50 W for all modulation types.
NOTE: Increments
The size of the increments depends on the intensity range setting.
From 1 to 10: Increments of 1
From 10 to 100: increments of 5
NOTE: Increments
The coarse adjustment increments are 1, 25, 50, 75, 100.
The size of the increments for fine adjustment depends on the intensity range
setting.
From 1 to 10: Increments of 1
From 10 to 100: increments of 5