4 Operation
Operating Instructions TENEO
Dentist element
ª
If the key is highlighted orange, the water tempering function is
activated.
2.
Use the – and + keys to set the water temperature.
Switching the instrument light on/off and adjusting the brightness
1.
Touch the
Instrument Lig
ht
key.
ª
If the key is highlighted orange, the instrument light on the SPRAYVIT
L multifunctional syringe is switched on, provided that it is the only
instrument removed from its holder.
If the SPRAYVIT L is not operated for 10 s, the instrument light
automatically switches off.
2.
Use the – and + keys to set the brightness.
Preselecting the instrument light operating voltage
The original Sirona halogen lamp is operated at 3.6 V. The operating voltage
can be set to max. 5 V in order to adapt it to different lamps.
Setting the operating voltage
¾
Use the – and + keys to adjust the operating voltage of the instrument
light.
4.5.12
SIROSONIC TL scaler
The SIROSONIC TL scaler is used to remove plaque and for endodontic
treatment.
4.5.12.1
Safety information
The torque wrench is used as a tool for screwing in instrument tips and, at the
same time, to protect against injury.
Attaching the torque wrench (SIROSONIC)
Only ultrasonic tips from Sirona
CAUTION: The operating voltages of different lamps may vary.
Overvoltages can lead to damage.
¾
When changing lamps, make sure that the operating voltage is properly
set for the new lamp.
NOTE:
Also observe the operating instructions of the SIROSONIC TL.
!
WARNING: Ultrasonic tips are sharply pointed.
There is a risk of injuring one's hand on the deposited scaler.
¾
Always attach the torque wrench to the scaler for protection as soon as
you deposit the handpiece.
WARNING: In some cases, ultrasonic tips from other
manufacturers do not guarantee safe operation.
Use only ultrasonic tips from Sirona.