61 93 556 D3509
69
D3509
.
201.01.02
.
02
19.09.2008
4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
2.
Use the – and + keys to set the flow rate of the NaCl pump.
3.
Touch the
Return
key.
ª
The
NaCl
settings screen is hidden immediately. The settings are
saved. If the
NaCl
key is highlighted orange, the function is activated.
Flow rate of the ultrasonic tip
4.5.9
Electric motor
4.5.9.1
Motor and coupling versions
Available electric motors include a collector motor (SL motor or SL motor with
ISO adapter), two brushless motors (BL or BL ISO motor), and a special
brushless implantology motor (BL Implant motor) for surgical interventions.
The SL motor and the motors of the BL line all require separate control
electronics. It is not possible to combine collector and brushless motors.
Collector motor
Collector motors are DC motors in which the voltage supply is realized via
carbon brushes. The speed range lies between 200 and approx. 40,000 rpm
(revolutions per minute).
SL motor
The
green
coded instrument hose (GN) must be used for the SL motor.
The SL motor is designed for direct operation of T1 Classic handpieces. In
order to connect e.g. T1 Line and T2 Revo handpieces, an ISO adapter must
be used.
CAUTION: Ultrasonic tips of third-party manufacturers in some
cases don't offer a sufficient flow rate in conjunction with the NaCl
function.
Use only ultrasonic tips from Sirona.
NOTE:
Also observe the operating instructions for the different motors.
GN