4 Operation
Sirona Dental Systems GmbH
Patient chair
Operating Instructions TENEO
Head pad magnet
Removing the MultiMotion headrest
1.
Press the release button
(A)
.
2.
Pull the complete headrest out of the headrest extension
(C)
.
3.
Cover the opening of the headrest extension with the cover cap
(B)
.
4.
If a children's head pad
(D)
is available, place it on the headrest
extension.
ª
The head pad is held magnetically.
Inserting the MultiMotion headrest
1.
Check the guide of the removed headrest for contamination and remove
any contamination you may find.
2.
Remove the cover cap
(B)
.
3.
Reinsert the MultiMotion in the guide from above until it audibly locks in
place. Pull on the headrest again to make sure that it is locked securely
in place.
4.4.8
Moving the patient chair via chair programs
The chair programs can be selected via the touchscreen or via the user
interface of the assistant element. The entry/exit and mouth rinsing positions
can also be selected via the foot control.
You can individually reprogram the factory preset chair programs to suit your
own wishes; see "Creating chair programs" [
.
You can select the chair program during which the dentist element is to
approach the treatment center; see "Linking dentist element movement to
chair programs" [
.
WARNING: The head pad contains a strong magnet on its bottom
side.
This magnet could affect any cardiac pacemaker located nearby.
Furthermore, direct contact of the head pad with magnetic cards can delete
data stored on the cards.
¾
Therefore, make sure that the magnet is never located in the immediate
vicinity of any patients, users or technical personnel wearing a cardiac
pacemaker.
¾
Make sure that no magnetic cards or any other data storage media are
located in the immediate vicinity of the head pad.
NOTE: Storage of the MultiMotion headrest
Store the MultiMotion in a safe place where it cannot fall onto the floor.