4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
4.5.9.4
Setting the direction of rotation
The direction of rotation can be changed only when the motor is at a
standstill.
Tip: After starting the electric motor with the foot switch, an audible
warning signal sounds 6 times if counterclockwise rotation is activated.
Via EasyPad
✔
An electric motor is removed from the holder.
➢ Press the
Counterclockwise rotation / User profile
key.
The
Counterclockwise rotation
status display lights up when
counterclockwise rotation is set.
Via the touchscreen of the EasyTouch
✔
An electric motor is removed from the holder.
✔
The
Motor dialog
is displayed on the touchscreen.
➢ Touch the
Counterclockwise rotation
key on the touchscreen.
For counterclockwise rotation: The key is highlighted orange.
For clockwise rotation: The key is displayed in gray.
Via C+ electronic foot switch
The rotational direction of the motor can also be set via the 4-way foot
control plate of the C+ electronic foot switch.
✔
An electric motor is removed from the holder.
1. Move the 4-way foot control plate to the left.
When counterclockwise rotation is switched on, the
Counterclockwise
status display lights up on the EasyPad, the
Counterclockwise
key is marked orange on the EasyTouch.
2. Move the 4-way foot control plate to the left again to reverse the
rotational direction.