4.11 Configurating the simulation unit (setup)
Operating Instructions Simulation Unit SIM MODULAR
4.11.1.4
Setting the purging time for the purge function
The purging time for the purge function may be set between 20 and 120
seconds.
✔
The setting focus
purge function
flashes.
1. Use favorite keys 1 and 3 to set the purge time.
2. Confirm with the
2 / Setup
favorite key.
4.11.1.5
Setting the purging time for the autopurge function
The purging time for the autopurge function can be set between 60 and
180 seconds.
✔
The setting focus
purge function
flashes.
1. Use favorite keys 1 and 3 to set the purge time.
2. Confirm with the
2 / Setup
favorite key.
4.11.1.6
Adjusting the cleaning agent mixture for chemical suction hose
cleaning
The chemical suction hose cleaning option is available for
SIM MODULAR.
The suction system can be automatically cleaned by pumping water into
a tank behind the receptacle of the suction hoses and extracting it from
there. A cleaning agent is added to the water if the simulation unit is
equipped with the chemical suction hose cleaning option. For more
information, see the section Suction hose cleaning with “suction hose
cleaning” option [ → 173].
It is possible to set how much cleaning agent should be added to the
water for chemical suction hose cleaning. The quantity is dependent on
the cleaning agent used and the type of treatment. Please follow the
manufacturer’s instructions for the cleaning agent.
✔
The setting focus
Cleaning agent mixture
flashes.
1. Use the favorite keys 1 and 3 to adjust the cleaning agent mixture for
chemical suction hose cleaning (0 to 5%).
2. Confirm with the
2 / Setup
favorite key.
4.11.1.7
Coupling suction to the 4-way foot switch.
If the simulation unit is equipped with a position selector valve for the
suction system, it can be set so suction flow can be interrupted or
restarted by moving the 4-way foot switch at the base of the chair in any
direction. Follow the safety instructions, see "Suction handpieces"
✔
The setting focus
Couple suction at 4-way foot switch
flashes.
1. Change the value with the favorite keys 1 or 3.
If a 1 is in setting focus, the suction can be turned on/off at the 4-
way foot switch.
2. Confirm with the
2 / Setup