4 Operation
Sirona Dental Systems GmbH
4.7 Stand-alone water supply
Operating Instructions Simulation Unit SIM MODULAR
Switching between the public drinking water supply and fresh water
bottle
If the simulation unit is equipped with the optional fresh water bottle in the
version which allows switching to the public drinking water supply, the
water supply can be switched by turning a lever.
1. For the water supply from the fresh water bottle, turn the lever
clockwise to the bottle symbol.
For operation with the public drinking water supply, turn the lever
counterclockwise to the faucet symbol.
2. Sanitize the water lines after switching to the public drinking water
supply (see Sanitizing with fresh water bottle).
Water supply, stand-alone, rinsing the water lines after switching to the fresh water bottle
CAUTION
After switching to the water supply from the fresh water bottle, there is
still water in the lines without added disinfectant.
➢ Rinse the water lines through for at least 30 seconds after switching.
This rinses out the water without added disinfectant from the
simulation unit and replaces it with the mixture of water and
disinfectant (see "Purge function on the EasyPad".