8 Electrical connection
Sirona Dental Systems GmbH
8.2 Connecting the line voltage
Installation Manual ORTHOPHOS XG 3D / Ceph
8.2
Connecting the line voltage
8.2.1
Connecting the unit
Connect the unit to the central distributor of the building installation,
preferably using a 3-wire power cable (N, L, PE, at least 3x2.5 mm² or 3x4
mm² (14 AWG or 12 AWG)).
25 A overcurrent circuit breaker B
.
IMPORTANT
When operating the unit at exhibitions and trade fairs, read the chapter
entitled Using demo mode – operation without radiation release [ → 173].
DANGER
Perilous shock hazard!
Switch off the power supply connection before connecting the line
voltage.
DANGER
Perilous shock hazard!
Fixed connection!
Installing a mains plug instead of the specified fixed connection infringes
international medical regulatory actions and is prohibited. In case of
error, this puts patients, users, and other parties seriously at risk.
A
Power cable to the unit
B
Second protective ground wire
C
Device
IMPORTANT
The unit is suitable for connection to networks of 200-240 V and 50-60
Hz, ±10%.
WARNING
Potentially lethal shock hazard!
Be sure to connect the second protective ground wire to ground.
L
N
PE
L
N
PE
PE
A
B
A
B
A
B
C