13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.2
3D adjustment/calibration
13.2.2.1
Sensor
1. Insert the geometry phantom in the pan bite block holder of the
2. Align the geometry phantom with the spirit level and tighten the screw
(A).
* For enhanced representation, the unit is illustrated without the
combisensor.
3. Call the
"Adjustment/calibration"
4. In the structure tree, under
"3D Adjustment/calibration"
, click on the
"Sensor "
element.
Once the menu has been selected via the button, the
"Sensor "
element (S010.20) is automatically pre-selected in the structure tree.
5. Make SIDEXIS XG ready for exposure. [ → 206]
6. Take an exposure (85 kV/23 mA s) [ → 206].
A message is displayed in the action window to indicate whether
adjustment has been successful and the traffic light symbol in
front of the
"Sensor "
element in the structure tree responds
accordingly:
Adjustment not OK = traffic light stays red
Adjustment OK = traffic light turns yellow
7. Repeat the procedure starting at Step 5 until adjustment is
successful.
8. When adjustment is OK, save the values.
*