11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.11 Using demo mode – operation without radiation release
båÖäáëÜ
11.11.3
Important information for repacking and transport
Cephalometer: XG3D / 3D ready / 5
Packing the cephalometer
1. Call service routine S034.6.
After calling the service routine, an inactive progress indicator is
displayed in selection field 1. Selection field 2 shows the
characters "FFFF".
2. Touch the T key (A).
The cephalometer moves to the packing position. The procedure
is visualized by an active progress indicator in selection field 1.
When the cephalometer has reached its packing position, "0000"
is displayed in selection field 2.
3. Exit the service routine [ → 129].
IMPORTANT
If a used carton on which one of the shockwatch or tiltwatch indicators
has already been tripped is used to package the unit, please make an
entry to that effect on the delivery note.