63 03 452 D3352
146
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.3.12
S017: Test step 17
Trigger save/restore function for DX61 data
* Factory setting
1. Call service routine S017.17 [ → 125].
Once the service routine has been selected, "FILTER" is
displayed in selection field 1.
2. Use the arrow keys (A) to select which type ("FILTER" or
"MECHANICS") you wish to check in selection field 1.
NOTICE! Check the need for adjustment for both "FILTER" and
"MECHANICS" one after the other.
3. Confirm your selection by pressing the T key (B).
Selection field
Parameter/Display
Meaning/Range of values
1
FILTER
Type of filter*
MECHANICS
Type of mechanics
2
For FILTER setting:
01
0.8 mm Cu (worldwide)
02
1.8mm Cu (BRD)*
03
Not in use
For MECHANICS setting:
01
ORTHOPHOS XG3D
ready
02
ORTHOPHOS XG 3D*
3
61 → 11
Direction of data transfer =
DX61 → DX11
11 → 61
Direction of data transfer =
DX11 → DX61
IMPORTANT
Filter type "02" (1.6 mm Cu) and mechanics type "02" (ORTHOPHOS
XG 3D) must be set in the ORTHOPHOS XG 3D volume tomograph.