11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.3.5
S017: Test step 6
Enable/disable the remote control
* Factory setting
S017_6 XG 3D / XG 3D ready
1. Call service routine S017.6 [ → 125].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the code for the required setting in
selection field 1 (see table).
Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
Selection field
Code
Function
1
00
Remote control
disabled*
01
Remote control
enabled
Содержание Orthophos XG 3D/Ceph