11 Initial startup
Sirona Dental Systems GmbH
11.6 Performing a mechanical function test
Installation Manual ORTHOPHOS XG 3D / Ceph
11.6
Performing a mechanical function test
The mechanical function of the rotation unit is checked via a test cycle.
The test cycle is executed without radiation. The test cycle is used to
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. The rotating unit stops automatically if the
resistance increases.
If any untypical noises occur during the test cycle or the rotation unit stops
due to increased resistance, the assembly of the combisensor must be
checked again.
Performing a 2D (panoramic) test cycle
✔
The unit must be switched on.
1. Touch the PAN symbol at the top of the touchscreen.
The exposure program for panoramic exposure (2D) is selected.
2. Press the T key.
On the touchscreen, the display of the kV/mA value, the
exposure time and the patient symbols is hidden. Two test cycle
symbols appear.
3. Press the release button.
The test cycle is started.
4. Wait until the test cycle has been completed.
T