59 38 399 D3352
2 – 8
D3352.076.01.13.02
07.2008
2.4 Error messages
Tab 2
2.4.2
Ex
yy
zz
Identifier
yy
defines the location or logical function unit where the error has
occurred.
The location may be a DX module number standing for an entire HW function
unit, or a logical SW function unit on board DX11 (central control).
2.4.3
Ex yy
zz
Identifier
zz
constitutes a consecutive number with the
error identification.
2.4.4
General handling of error messages
Error messages always must be acknowledged with the
R key
.
If failure-free operation is possible after the error is acknowledged, then no
further action is necessary.
If error messages reoccur or occur frequently, identify the error as described
in section 2.5 and take appropriate action to eliminate the corresponding error
or fault.
In some cases, it may make sense to obtain more information on the history
and frequency of errors via the error logging memory (S007) and
SIXAB
-
CON
‡
PROPERTIES
‡
EXTENDED
DETAILS
(see section 1.10 on page 1-32)
(see also section 3.1).
yy
Location/Function unit
Board
06
Tube assembly
DX6
61
Diaphragm control (not for XG 3/3 PPE, fixed diaphragm)
DX61
07
User interface on the Easypad (XG
Plus
)
DX7
71
User interface on the Multipad (XG5/3/3 PPE)
DX71
10
System hardware
DX11/DX1
11
System software
DX11/DX1
12
CAN bus
DX11/DX1
13
Stand peripherals
DX11/DX1
14
Digital extension
DX11/DX1
15
Configuration/update (wrong software, wrong module constellation, etc...)
DX11/DX1
41
Media interface card (omitted as of HW version BA, November 2006)
DX41
42
Remote control
DX42
81
Sensor (Pan slot: zz=1-23 ; ceph slot: zz= 51-73)
DX81
91
Cephalometer
DX91
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...