2.1 Help messages
Tab 2
2.1
Help messages
Help messages are displayed as help codes (Hxxx) on the touchscreen of the
Easypad (XG
Plus
), on the display of the Multipad (XG5/3/3 PPE) or on the
remote control display (if available). The codes tell you how to operate the
system if radiation release is not possible due to a previous operator error.
The following list provides you with an overview of all help codes, their mean-
ing and the action required to eliminate the corresponding problems:
Help code
Description
Actions required
H3 01
The rotating element on the panoramic unit is
not located in its starting position.
z
Press the R key:
The panoramic unit moves to the starting
position.
H3 10
The system is performing an action.
z
Wait until the system is ready.
H3 20
The exposure parameters have not been
acknowledged yet.
z
Press the R key:
The exposure parameters are confirmed.
H3 21
The X-ray room door contact is not detected.
z
Close the door or check door contact.
H3 22
(for XG
Plus
only)
Quadrant for TSA exposure is not selected.
z
Select quadrant via Easypad.
H3 30
(XG 3 PPE only)
Indicates number of exposures remaining
(PPE concept).
1. After each switch-on (see S017.22)
2. After each exposure if the number of expo-
sures remaining is below the defined warning
level (see S017.23)
z
Request new activation code via Sirona
Customer Service Center (CSC) in due
time.
H3 31
(XG 3 PPE only)
No further exposures available.
z
Enter new activation code.
H4 01
The sensor on the panoramic unit is not properly
plugged in or is missing.
z
Plug sensor into PAN slot.
HINWEIS
i
If this message does not disappear after the
sensor is plugged in, this indicates a system er-
ror. Perform error diagnosis according to
Section 3.2.
H4 02
The sensor on the cephalometer is not properly
plugged in or is missing.
z
Plug sensor into Ceph slot.
HINWEIS
i
If this message does not disappear after the
sensor is plugged in, this indicates a system er-
ror. Perform error diagnosis according to
Section 3.2.
H4 03
SIDEXIS is not ready for exposure.
z
Make SIDEXIS ready for exposure.
H4 04
The sensor does not match the selected expo-
sure type
z
Plug in the ceph sensor.
H4 05
The sensor does not match the selected expo-
sure type
z
Plug in the TSA sensor.
H4 06
The cephalometer is not located in its starting
position.
z
Press the R key:
The cephalometer moves to the starting po-
sition.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...