båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 145
Tab 5
5.19 Service routine S034
5.19
S034.7: Display/change ring type
5.19.4
S034: Test step 7
(XG
Plus
only)
Selecting test step 7
1.
Select test step 7 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
2.
Select the ring type using the arrow keys of selection field 1:
01 = old ring type
= PAN unit serial numbers up to 20056
PAN/CEPH units up to serial number 40104
02 = new ring type
= PAN units with serial no. 20056 and higher
PAN/CEPH units with serial no. 40104 and higher
3.
To save the setting, first press the
Memory key
(R key lights up)
and then press the
R key
.
1
2
3
+
-
S034
7
XG
Plus
:
Easypad touchscreen
+
-
(
o
S34
(
n
1
2
3
+
-
01
XG
Plus
:
Easypad touchscreen
R
(
n
S34
R
R
1
2
3
+
-
01
XG
Plus
:
Easypad touchscreen
R
)
n
)
o
S34
7
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...