59 38 399 D3352
4 – 32
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
4.3.5
Adjusting the (fixed) pan diaphragm
(XG 3 / 3 PPE)
Menu: Diaphragm/system adjustment
‡
Pan - Diaphragm
Preparations
NOTE
i
The fixed diaphragm must be reattached for this adjustment step
(see page 4-10).
z
Remove the needle phantom from the bite block holder of the panoramic
X-ray unit (see page 4-9).
Opening
PAN
-
DIAPHRAGM
1.
Go to the
PAN
-
DIAPHRAGM
submenu.