7 Installation: Remote control

Sirona Dental Systems GmbH

7.4 Connecting the control cables (REMOTE)

Installation Manual ORTHOPHOS SL

7.4

Connecting the control cables (REMOTE)

7.4.1

Installation version 1: Without coiled cable

1. Unscrew clamp (A) from the board.
2. Place cable L117 in the clamp so that the turned up cable shield is 

completely enclosed.

3. Re-attach the clamp to the board.
4. Connect control cable L117 to terminal X108 (board DX42) as shown 

in the connection diagram. Shorten control cable L117 as required.

5. Lay cable L117 in a loop following the bottom edge of the chassis.
6. Close the housing.
For information on how to connect a door contact switch, refer to the 

section entitled “Connecting a door contact switch [ → 80]”.

1
2
3
4
5
6

XRAY
XRAY

GND

+28 V

CAN-H

CAN-L

Pin

1
2
3
4
5
6

15 m
590“

L117

X108

X108

L117

DX42

WHGN
GN
BN
WHBN
BU/WHOG
WHBU/OG

PIN

Code

Color

1

WHGN

White/green

2

GN

Green

3

BN

Brown

4

WHBN

White/brown

5

BU/WHOG

Blue / white/orange

6

WHBU/OG

White/blue / orange

A

Содержание ORTHOPHOS SL 2D

Страница 1: ...k W f j MPKOMNS loqelmelp pi Oa loqelmelp pi Oa L loqelmelp pi Pa loqelmelp pi Pa L b b Cover page...

Страница 2: ......

Страница 3: ...ns of service personnel 13 2 6 Switching the unit on 13 2 7 Radiation protection 13 2 8 Laser light localizer 13 2 9 Modifications to the unit 14 2 10 Transport safety devices 14 2 11 Electromagnetic...

Страница 4: ...1 Standard version 42 5 1 2 Option 1 with second wall holder 44 5 1 3 Option 2 Floor stand installation 45 5 2 Tools materials and measurement tools you will need 47 5 2 1 Tools and materials 47 5 2 2...

Страница 5: ...Installation Ceph arm 89 10 1 Installation material 89 10 2 Required tools 90 10 3 Installation 91 10 3 1 Ceph arm mounted on left hand side 91 10 3 2 Ceph arm mounted on right hand side 94 10 3 3 In...

Страница 6: ...lect parameters 135 11 12 2 4Saving parameters 135 11 12 2 5Exiting the test step and service routine 136 11 12 3 S017 Configuration service 137 11 12 3 1S017 Test step 2 138 11 12 3 2S017 Test step 3...

Страница 7: ...calibration 193 13 1 1 Displays and help messages during adjustment and calibration 194 13 1 2 Calibration menu 195 13 1 2 1 Opening the calibration menu 195 13 1 2 2 Menu structure 197 13 1 3 Enabli...

Страница 8: ...Diaphragm adjustment 264 13 2 2 3 Sensor calibration 265 13 2 2 4 Diaphragm configuration 266 13 2 2 5 Geometry calibration 268 13 2 2 6 Dosimetry 270 13 2 3 Resetting adjustment calibration 271 13 3...

Страница 9: ...ety information provided in the present operating instructions Such information is highlighted as follows Tip Information on making work easier DANGER An imminent danger that could result in serious b...

Страница 10: ...ed The formats and symbols used in this document have the following meaning Prerequisite 1 First action step 2 Second action step or Alternative action Result Individual action step Requests you to do...

Страница 11: ...er relevant documents In addition to these installation instructions you will require the following documentation Orthophos SL wiring diagrams Wiring diagrams ORTHOPHOS SL Wiring References REF 64 95...

Страница 12: ...sleeves Identification of single use devices Prior to each exposure the hygienic protective sleeves single use devices must be fitted Single use devices are identified with the symbol shown on the lef...

Страница 13: ...the X ray tube assembly as the coiled cable of the manual release permits With the exception of the service engineer no other persons are allowed to stay in the room during an exposure In case of mal...

Страница 14: ...edical electrical equipment is subject to special EMC related precautions It must be installed and operated as specified in the document Installation Requirements If high voltage systems radio link sy...

Страница 15: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 15 Sirona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 1 System version b 3 Unit description 3 1 System version A B C...

Страница 16: ...THOPHOS SL 3D Digital X ray unit with digital Flat Panel Detector and DCS sensor Flat Panel Detector Digital Flat Panel Detector with a Si technology amorphous silicon Pan sensor Digital CdTe sensor w...

Страница 17: ...ion Manual ORTHOPHOS SL 3 2 Sensor versions b 3 2 Sensor versions 2D sensor unit Sensor unit for panoramic X ray PAN DCS sensor 2D 2D 3D sensor unit Sensor unit for panoramic X ray PAN and volume expo...

Страница 18: ...31 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements 3 3 1 Top view A Recommended dist...

Страница 19: ...6 19 Sirona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 2 Front view A Alternative fastening if it is not possible to screw the unit...

Страница 20: ...D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 3 Top view with floor stand A Recommended distances fro...

Страница 21: ...ona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 4 Front view with floor stand A Floor stand Order floor stand separately The unit mu...

Страница 22: ...2 D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 5 Top view with Ceph left A Recommended distances from...

Страница 23: ...rona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 6 Front view with Ceph left A Alternative fastening if it is not possible to screw...

Страница 24: ...4 D3632 031 01 02 02 03 2016 3 Unit description Sirona Dental Systems GmbH 3 3 Dimensions Space requirements Installation Manual ORTHOPHOS SL 3 3 7 Top view with Ceph right A Recommended distances fro...

Страница 25: ...ona Dental Systems GmbH 3 Unit description Installation Manual ORTHOPHOS SL 3 3 Dimensions Space requirements b 3 3 8 Front view with Ceph right A Alternative fastening if it is not possible to screw...

Страница 26: ...section Wall mounting standard and option 1 49 B Option 1 Wall mounted installation with 2 wall holders short and no floor fastening if only wall installation is possible on site see section Wall moun...

Страница 27: ...mote control with release button on the coiled cable in the treatment room B Installation version 1 Unit with remote control outside the X ray room without release button on the coiled cable see secti...

Страница 28: ...are attached to the side of the X ray unit packaging to indicate whether the unit was exposed to a shock during transport White indicator No shock Red indicator Shock Transport and storage temperatur...

Страница 29: ...t doesn t necessarily mean that the unit is damaged Make a note on the delivery slip that the indicator is activated Ensure that the driver from the transport company confirms this on the delivery sli...

Страница 30: ...designed both for protection during transport and as an installation aid Therefore please remove only the surrounding packaging prior to installation Please leave the Styrofoam packaging and transpor...

Страница 31: ...e save the two lateral Styrofoam packaging parts Scope of supply B Filling piece 1 C Filling cardboard 1 D Pack list 1 E Filling cardboard 2 F Filling piece 2 including pack list 2 G Profile cover H I...

Страница 32: ...D3632 32 D3632 031 01 02 02 03 2016 4 Delivery and transport Sirona Dental Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL 4 2 2 Sensors DCS sensor A Styrofoam packaging Scope of supply B D...

Страница 33: ...D3632 031 01 02 02 03 2016 33 Sirona Dental Systems GmbH 4 Delivery and transport Installation Manual ORTHOPHOS SL 4 2 Delivery b Flat Panel Detector A Styrofoam packaging Scope of supply B Flat Pane...

Страница 34: ...0 3 4 x 28 3 4 Weight kg 40 lbs 88 IMPORTANT The right handed ceph arm is packed laterally reversed A Styrofoam packaging Scope of supply B Ceph arm C Accessories 35 and hygienic protective sleeves 35...

Страница 35: ...ion Manual ORTHOPHOS SL 4 2 Delivery b 4 2 3 1 Accessories 4 2 3 2 Hygienic protection 20 30 40 50 C A B A Nose support 1x B Ear plug holders with ear plug fixation 2x C Carpus support plate 1x 20 30...

Страница 36: ...Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL 4 2 4 Adjustment sets 2D adjustment Panoramic 2D adjustment Cephalometer A A Needle phantom B B Set of Torx offset screwdrivers A C B A Ceph...

Страница 37: ...Dental Systems GmbH 4 Delivery and transport Installation Manual ORTHOPHOS SL 4 2 Delivery b 3D adjustment calibration Acceptance and constancy test Worldwide 3D constancy test worldwide A A Geometry...

Страница 38: ...02 02 03 2016 4 Delivery and transport Sirona Dental Systems GmbH 4 2 Delivery Installation Manual ORTHOPHOS SL For Germany only 3D constancy test Germany B A A DVT test phantom DIN B Contrast element...

Страница 39: ...ackaging attached normal case 1 Open the surrounding packaging using the straps provided A 2 Pull out the carrying handles B 3 Transport the unit to the installation site NOTICE Possible transport dam...

Страница 40: ...ral styrofoam packaging parts 4 Loosen the four screws B 5 Pull the carrying handles C out of their holders and insert them through the drillings D of the wooden support A from the inside 6 Insert scr...

Страница 41: ...d transport it to the installation site NOTICE Possible transport damage The pan sensor and flat panel detector are sensitive instruments Only remove the packaging immediately prior to installing the...

Страница 42: ...016 5 Installation X ray unit Sirona Dental Systems GmbH 5 1 Installation material Installation Manual ORTHOPHOS SL 5 Installation X ray unit 5 1 Installation material 5 1 1 Standard version A For ins...

Страница 43: ...all floor mounting Hexagon wood screw 8x80 5 16x3 4 pc Plastic wall plug S10 2 pc Screw M8x30 2 pc Washer 8 4 6 pc M8 nut 2 pc Screw M4x10 18 pc Washer 4 3 2 pc Mounting kit 10 SXR 2 pc TORX offset sc...

Страница 44: ...al ORTHOPHOS SL 5 1 2 Option 1 with second wall holder Additional wall holder for bottom wall mounting A For installation on wooden stud frame structures Wall holder long 1 pc Hexagon wood screw 8x80...

Страница 45: ...2 02 03 2016 45 Sirona Dental Systems GmbH 5 Installation X ray unit Installation Manual ORTHOPHOS SL 5 1 Installation material b 5 1 3 Option 2 Floor stand installation A Mounting hardware B Floor st...

Страница 46: ...al ORTHOPHOS SL Installation with floor stand Floor stand 1 pc Floor stand cover 1 pc Wood screw 10x160 3 8x6 5 pc Plastic wall plug S12 5 pc Screw M8x80 2 pc Washer 8 4 2 pc M8 nut 2 pc Screw M10x50...

Страница 47: ...10mm 3 8 Drill hammer drill general purpose Drill or drill hammer depending on the ground Graphics Open end wrench neutral Open end wrench 13 mm A F Open end wrench 13 mm A F Socket wrench Wrench inse...

Страница 48: ...wrench Hexagon socket 2 mm Allen key 2 mm Socket wrench Socket wrench and extension 17 Socket wrench and extension Wrench insert 17 5 2 2 Measurement tools Multimeter or ammeter Test unit to measure u...

Страница 49: ...cover A 2 Position the two installation aids B at the foot C of the unit and secure their position with adhesive tape CAUTION The installation aids must be placed on top of each other in such a way t...

Страница 50: ...thout a pallet set the unit upright with the wooden support You can also use the lateral styrofoam packaging as a support with this variation 4 Loosen the nuts with washers D on both sides of the pall...

Страница 51: ...llation X ray unit Installation Manual ORTHOPHOS SL 5 3 Wall mounting standard and option 1 b Mounting the first upper wall holder standard Mount the upper wall holder Mounting the second lower wall h...

Страница 52: ...holder must be flush with the upper edge of the unit The unit is now mounted on the upper wall holder 4 Insert screw C through washer D and then through the wall holder and into the stand from behind...

Страница 53: ...L 5 3 Wall mounting standard and option 1 b Leveling the unit and fastening it to the floor standard and option 1 1 Remove the transport safety device G 2 Pull wooden board H out of the styrofoam pack...

Страница 54: ...lp of the spirit level by adjusting the unit base Align the stand by placing the spirit level on the side and rear of the stand I Check the ring in both directions using the spirit level J 5 Drill thr...

Страница 55: ...using the floor stand it must be secured with the upper wall holder NOTICE Reduced image quality Note the wall and ground conditions If a carpet is present at the unit installation site it must be re...

Страница 56: ...packaging the two lateral styrofoam parts and the profile cover A 3 Loosen the nuts with washers B on both sides of the pallet or wooden support 4 Take off the pallet or wooden support 5 Remove the th...

Страница 57: ...k G NOTICE Do not damage the interfaces or cables Make sure that the interfaces and cables do not bear on the supporting block G and are not damaged when you push the unit 10 Remove the two screws H f...

Страница 58: ...L firmly onto the base plate N using the 4 screws and washers M 13 NOTICE Do not damage the cables Ensure that the cables are correctly routed through the support and are not crushed Position the bas...

Страница 59: ...ty device 63 2 Insert screw P through the washer and then from the rear through the support and into the stand 3 Fit the profile clamp Q onto the screw P from the other front side and screw the nut R...

Страница 60: ...unterclockwise to the front side of the unit 7 Loosen screws P and adjusting nuts O again slightly 8 Level the unit in both directions with the help of the spirit level by adjusting the adjusting nuts...

Страница 61: ...3 Attach the unit loosely to the upper wall holder Place the screws U in the grooves Screw the unit onto the wall holder loosely using nuts and washers V NOTICE Do not tighten the screws firmly 4 Leve...

Страница 62: ...r stand option 2 Installation Manual ORTHOPHOS SL 5 Drill the fastening holes in the floor through the holes in the base plate Remove the drilling dust with a vacuum cleaner 6 Slide the wall plugs W t...

Страница 63: ...in a safe place You will need them should the unit be moved again 5 6 Installing tube assembly cover 1 Loosen screw A on the unit s swivel approx 3 4 turns 2 Fit the Top tube assembly cover B under th...

Страница 64: ...holder Installation Manual ORTHOPHOS SL 5 7 Installing the release button holder 1 Unscrew and remove the cover A 2 Attach the holder B 3 Screw the cover A onto the unit B A IMPORTANT Only attach the...

Страница 65: ...tools Screwdriver Torx TX10 Screwdriver Torx T10 Screwdriver Torx TX20 Screwdriver Torx TX20 NOTICE Damage to the Flat Panel Detector The Flat Panel Detector is a sensitive component Be careful not t...

Страница 66: ...er on the ring Installation Manual ORTHOPHOS SL 6 2 Installing the sensor cover on the ring 1 Loosen screw G approx 2 3 turns 2 Insert the sensor cover H from below over the sensor bracket and then sl...

Страница 67: ...lation Manual ORTHOPHOS SL 6 3 Attaching DCS sensor b 6 3 Attaching DCS sensor 1 Carefully remove the DCS sensor from the packaging 2 Fit the DCS sensor into the holes A of the rotation unit and secur...

Страница 68: ...g the Flat Panel Detector 1 Carefully remove the Flat Panel Detector from the packaging 2 Fit the Flat Panel Detector from above into the guide of the rotation unit and secure the detector immediately...

Страница 69: ...5 Connecting a sensor unit b 6 5 Connecting a sensor unit 1 Insert leads 1 L83A and 2 L83B from the rotation unit as shown in the illustration into plugs X100 X700 and X800 on board DX83 2 If a Flat P...

Страница 70: ...5 142 D3632 70 D3632 031 01 02 02 03 2016 6 Installing the sensor unit Sirona Dental Systems GmbH 6 5 Connecting a sensor unit Installation Manual ORTHOPHOS SL 3 Lay the leads as shown in the illustra...

Страница 71: ...GmbH 6 Installing the sensor unit Installation Manual ORTHOPHOS SL 6 6 Final installation work b 6 6 Final installation work Connect the sensor covers D and E together over the sensor and secure them...

Страница 72: ...ll or drill hammer depending on the ground Graphics Slot screwdrivers assorted Slot screwdriver neutral size Slot screwdriver Torx offset screwdriver graphics Torx offset screwdriver TX20 Torx offset...

Страница 73: ...chassis from underneath Preparations 1 Press carefully into the groove A from below with the slotted screwdriver do not lever 2 Remove the lid from the chassis 3 Hold the chassis against the wall in...

Страница 74: ...horten cable L117 to the desired length 2 Expose the cable shield D see condition on delivery The length of the wires should be approx 60 mm 2 3 8 3 Place the shortened shielding over the insulation a...

Страница 75: ...mote control Installation Manual ORTHOPHOS SL 7 3 Installation b Attaching the strain relief Left concealed installation right surface installation Attach cable L117 to the strain relief in the chassi...

Страница 76: ...to the board 4 Connect control cable L117 to terminal X108 board DX42 as shown in the connection diagram Shorten control cable L117 as required 5 Lay cable L117 in a loop following the bottom edge of...

Страница 77: ...the rear 2 Using the two screws C fasten the release button holder B to the keyboard 3 Plug the connector of the coiled cable into socket X101 on board DX42 and screw the connector down tight 4 Hook...

Страница 78: ...surface installation 2 Attach the cable to the still exposed strain relief 3 Connect the current carrying cables and the ground wire to terminal X105 board DX42 as shown in the connection diagram 4 Se...

Страница 79: ...ORTHOPHOS SL 7 6 Final installation work b 7 6 Final installation work 1 Reassemble the remote control 2 NOTICE Make sure that no cables are pulled off when you clip the cover on Hang the release but...

Страница 80: ...S SL 7 7 Connecting a door contact switch Connect the door contact switch between terminal X108 pin 1 board DX42 and control cable L117 pin 1 BU X108 1 2 3 4 5 6 Pin WHGN GN BN WHBN BU WHOG WHBU OG X1...

Страница 81: ...on version without remote control Connect the coiled cable of the release button to socket X103 of the unit Installation version with remote control 1 Use cable L117 to connect the remote control to s...

Страница 82: ...ing demo mode operation without radiation release DANGER Perilous shock hazard Switch off the power supply connection before connecting the line voltage DANGER Perilous shock hazard Fixed connection I...

Страница 83: ...leakage current according to IEC 62353 2007 see sections Checking the protective ground wires 84 and Checking the leakage current of the unit 88 2 Record the measured values in the document titled Ins...

Страница 84: ...on 2 Disconnect the power cable and the second protective ground wire from the building installation 3 Remove the bottom profile front tube assembly and rear tube assembly covers Measuring setup for p...

Страница 85: ...ng the formula R U I If the resistance value is greater than indicated in the adjacent table check whether the protective ground wires are fastened according to the specifications Check whether plain...

Страница 86: ...64 95 142 D3632 86 D3632 031 01 02 02 03 2016 9 Safety checks Sirona Dental Systems GmbH 9 1 Checking the protective ground wires Installation Manual ORTHOPHOS SL Measuring point A central ground wire...

Страница 87: ...ORTHOPHOS SL 9 1 Checking the protective ground wires b Measuring points B and C GNYE power connection and 2nd protective ground wire Measuring points D and E Board cage DX32 and tube assembly housing...

Страница 88: ...instructions of the tester 7 Document the measured value of the leakage current in the technical document Inspection and maintenance and safety related checks REF 64 95 100 to identify changes from th...

Страница 89: ...h arm 10 1 Installation material NOTICE Reduced image quality For operation of the unit with floor stand and Ceph an additional mounting using the upper wall holder is absolutely necessary No Ceph ins...

Страница 90: ...ols Graphics Open end wrench neutral Open end wrench 8 mm A F Open end wrench 8 mm A F Open end wrench 10 mm A F Open end wrench 10 mm A F Graphics Neutral Allen wrench Allen wrench 2 5 Allen wrench 2...

Страница 91: ...A 2 CAUTION Do not yank out the ambient light cable Remove the cover B Tip You can even leave the cover B on the unit and push it upwards a little 3 Remove the cover C by pulling it upwards 4 Reattach...

Страница 92: ...ofoam packaging on top 3 Remove the adhesive tape and the cover E from the ceph packaging 4 Break off the wall part F of the packaging at the perforation 5 Slide the cardboard support with the ceph ar...

Страница 93: ...ame height as the drill holes on the ceph support arm NOTICE Check the room height before you raise the unit If the room height is less than 2 27 m 89 3 8 or 2 30 m 90 1 2 for installation with the fl...

Страница 94: ...t 1 Loosen the screws A 2 CAUTION Do not yank out the ambient light cable Remove the cover B Tip You can even leave the cover B on the unit and push it upwards a little 3 Remove the cover C by pulling...

Страница 95: ...o the wall so that the drill holes of the support arm are horizontally aligned with the positions of the threaded bolts 6 Switch on the unit G During the unit startup phase the start screen appears on...

Страница 96: ...keys I on the control panel to lower the unit until the bearing bolts on the unit are at the same height as the drill holes on the ceph support arm NOTICE Check the room height before you raise the un...

Страница 97: ...n them firmly 3 Detach the cables from the holder Lay the cables over the support arm in the groove between the slide and the support arm from left to right toward the opening K 4 NOTICE Never crush t...

Страница 98: ...ragm 1 Remove the secondary diaphragm from the packaging 2 Remove the safety cable tie A for the screws 3 Guide the secondary diaphragm beneath the ceph arm and slide the connector upward through the...

Страница 99: ...2 Screw in place the plug connections L35 L38 3 Roll the plug connection L36 L39 into a loop and secure it with a cable tie 4 Grasping inside the ceph arm tube from below carefully pull cable L35 two...

Страница 100: ...02 02 03 2016 10 Installation Ceph arm Sirona Dental Systems GmbH 10 4 Connecting control cables Installation Manual ORTHOPHOS SL 5 Roll the cable into a loop and stow it away in the tube 6 Close the...

Страница 101: ...uide the cable on the unit through the groove A 2 Lay the cable and the cable loop L36 L39 as illustrated 3 NOTICE Ensure that you do not crush or kink the cables Attach the cover B 10 4 2 Running the...

Страница 102: ...the secondary diaphragm 2 NOTICE Do not damage the unit Do not move the secondary diaphragm or sensor adapter by hand or apply any force Remove the top cover 3 Remove the protective canvas A 4 Loosen...

Страница 103: ...ystems GmbH 10 Installation Ceph arm Installation Manual ORTHOPHOS SL 10 5 Final installation work b 8 Insert both ear plug holders and fasten them securely using the screws E 9 Press the button F and...

Страница 104: ...ers 1 For installation with floor stand only Place the four spring clips E on the base plate The two covers F and G must hold together after they have been attached 2 Attach the profile cover J To do...

Страница 105: ...ensor for ceph versions NOTICE Ensure that the correct sensor for cephalometry ceph sensor is plugged in Insert the ceph sensor A carefully upward into the holder until it audibly clicks into place 11...

Страница 106: ...e when it moves exception Calibration block and positioning aids NOTICE Fluctuations in temperature can cause condensation to form in the unit Electrical components are destroyed by short circuits Do...

Страница 107: ...lights up This LED remains lit as long as the unit is on The start screen is displayed on the touchscreen for several seconds The program selection is then displayed on the touchscreen The forehead su...

Страница 108: ...ration on delivery The acoustic signal for end of exposure is activated English is preconfigured as the unit language The preview image is switched on The welcome screen is switched on The first name...

Страница 109: ...e test cycle or the rotation unit stops due to increased resistance the assembly of the sensor must be checked again 11 6 1 Starting the test cycle The test cycle is executed without radiation The tes...

Страница 110: ...move the unit up and down through its entire adjustment range 2 Check the height adjustment mechanics Does the height adjustment function correctly Are there any unusual noises during the adjustment...

Страница 111: ...service routine S018 2 Unit variants For unit orders with the ceph arm the unit version must be configured through service routine S017 2 138 Remote control All units ordered with or without remote co...

Страница 112: ...ssigned a unique IP address and the name of the X ray component must also be unique Factory setting of the TCP IP address of the unit 192 168 15 240 subnet mask 255 255 255 0 NOTE The network configur...

Страница 113: ...an X ray component SIDEXIS 4 is installed The Sirona AcquisitionServer software is installed The workstation plug in software for the corresponding X ray component or sensor is installed on the workst...

Страница 114: ...ray components 4 If necessary Select the acquisition server from the Acquisition Server list field 5 Click on the Next button A password dialog box appears 6 Enter the service password and confirm you...

Страница 115: ...available X ray components The X ray components found on the network are displayed in list form NOTE If no X ray component is found check the availability of the unit and repeat the Automatic search...

Страница 116: ...ice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 6 Click on the Cancel button 7 Reboot the SIDEXIS 4 8 Register a patient...

Страница 117: ...ens Boot service 1 Click the Boot service button in the structure tree 2 A message window prompts you to switch the new X ray component back on to reboot The network is searched for the booting X ray...

Страница 118: ...st the network configuration IP address subnet mask and standard gateway to your network 4 Click the Save button to save the settings 5 Wait until the saving operation has been completed NOTICE If you...

Страница 119: ...vice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 7 Click on the Cancel button 8 Reboot the SIDEXIS 4 9 Register a patient...

Страница 120: ...anually The menu for adding X ray components opens Add device manually 1 Click the Add device manually button in the structure tree 2 Enter the IP address and the location of the desired X ray compone...

Страница 121: ...vice Administration The X ray component appears in the device list of the configuration menu ORTHOPHOS SL Device Administration 6 Click on the Cancel button 7 Reboot the SIDEXIS 4 8 Register a patient...

Страница 122: ...mo mode and repacking Sirona Dental Systems GmbH 11 9 Checking the data paths Installation Manual ORTHOPHOS SL 11 9 Checking the data paths Checking the 2D and 3D data paths is possible through the se...

Страница 123: ...ernet Explorer or Firefox 2 In the address line enter https and the IP address of your unit Example https 192 168 15 240 NOTICE The IP address of your unit can be found on the Info screen Calling the...

Страница 124: ...na Dental Systems GmbH 11 10 Setting ambient light via the Web service Installation Manual ORTHOPHOS SL 5 Change the color by clicking the color scale and brightness by clicking on the brightness scal...

Страница 125: ...Set dip switch S2 DX6 to position 2 IMPORTANT If switch S2 is not set to position 2 in demo mode before switching off the unit various error messages will display when the unit is turned back on 5 Pu...

Страница 126: ...connector J6 J103 DX6 Radiation release is now once again possible 6 Switch on the unit and check the mode on the info screen Demo mode ON means that Demo mode is switched on radiation release is not...

Страница 127: ...repacking and transport Packing the ORTHOPHOS SL IMPORTANT If a used carton on which one of the shockwatch or tiltwatch indicators has already been tripped is used to package the unit please make an...

Страница 128: ...assembly guard supplied in the package to the X ray tube assembly and secure the packing position of the tube assembly with the tension belt supplied in the package 5 Attach the transport safety devi...

Страница 129: ...utines S008 3 S017 2 S017 18 and S037 4 can also be run via the Sirona web service see section Running service routines via the web service 174 Service routines S017 and S037 must sometimes be perform...

Страница 130: ...OS SL 11 12 1 Service menu and service routines You can use the service routines to check the function of certain unit components and modules as well as to set important unit parameters 11 12 1 1 Disp...

Страница 131: ...and arrow keys to select unit parameters in the selection fields F S1 S37 Selected service routine G 1 n Selected test step H Patient symbol keys Different functions depending on service routine I R...

Страница 132: ...ines can be activated and unit settings incl tests and calibrations can be performed in this menu The touchscreen display is located on level 1 1 Touch the toothed wheel A in the upper right corner of...

Страница 133: ...a the arrow keys in the selection field 1 A and confirm the selection via the service key B If the selected service routine has several test steps the first selectable test step is displayed in select...

Страница 134: ...prevents the inadvertent selection or activation of these service routines To select a service routine or test step with security access proceed as follows 1 Select the service routine or the test ste...

Страница 135: ...ce routine S017 6 the code 01 is pre selected for the Remote control enabled option Press the or arrow keys A to select code 00 B for the Remote control disabled option Once the selected parameter has...

Страница 136: ...12 Service routines for commissioning and repacking Installation Manual ORTHOPHOS SL 11 12 2 5 Exiting the test step and service routine Touch the Service key A or the double arrow key B to go back to...

Страница 137: ...l s touchscreen Additional service routines can be performed Before a patient exposure can be performed the error message must be acknowledged with the R button and the unit must be restarted SR SA Fu...

Страница 138: ...on field 1 2 Use arrow keys A to select the required unit variants in selection field 1 see table Once the unit variant has been selected the Memory key B lights up 3 Save the setting 4 Exit the servi...

Страница 139: ...e routines for commissioning and repacking b Checking the jumper position on board DX1 Check the jumper positions of the sockets on board DX1 X309 DX91 ceph sensor and X503 DX81 ceph sensor Without ce...

Страница 140: ...country group code Factory setting 1 Call service routine S017 3 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to se...

Страница 141: ...4 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use the arrow keys A to select the code for the required language in selection field 1...

Страница 142: ...e control display factory settings 1 Call service routine S017 6 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use the arrow keys A to...

Страница 143: ...y setting 1 Call service routine S017 8 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select the required setting...

Страница 144: ...come screen Factory setting 1 Call service routine S017 13 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select th...

Страница 145: ...de for the line currently selected is displayed in selection field 1 2 Use arrow keys A to select the required line in selection field 1 see table The activation status code is displayed in selection...

Страница 146: ...re 1 Call service routine S017 15 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select the required setting in sel...

Страница 147: ...l angle deviating from the factory setting 7 make the corresponding configuration settings via service routine S018 7 factory settings 1 Call service routine S017 18 147 Once the service routine has b...

Страница 148: ...tment S018 2 No Set the maximum travel height S018 3 No Undo the maximum travel height setting S018 4 No Check the height adjustment sensor system S018 5 No Setting the minimum travel height S018 6 No...

Страница 149: ...user mode on the control panel 2 Call service routine S018 2 Once the service routine has been selected the current height position is displayed in selection field 1 The Memory key B lights up 3 To s...

Страница 150: ...RTHOPHOS SL 11 12 4 2 S018 Test step 3 Undo the maximum travel height setting 1 Call service routine S018 3 Once the service routine has been selected the current height position is displayed in selec...

Страница 151: ...een selected the current height position is displayed in selection field 1 Patient symbol keys 1 to 3 F show the switching state of the limit switches see table If the RESPECTIVE patient symbol key is...

Страница 152: ...n selected the current height position is displayed in selection field 1 The Memory key B lights up 3 To save the minimum travel height touch the Memory key B R key C lights up followed by the R key C...

Страница 153: ...e nominal value displayed in selection field 2 applies The nominal value for the program group selected in selection field 1 is displayed in selection field 2 The currently measured angle of the bite...

Страница 154: ...ible range 17 to 22 Tip You can also set the required angle manually on the occlusal bite block and transfer this angle to selection field 2 by pressing the T key J By pressing patient symbol key 1 G...

Страница 155: ...splayed in selection field 1 The currently measured AD value leading to the angle calculation is displayed in selection field 2 The codes of the possible error modes of the acquisition electronics are...

Страница 156: ...onnections on board DX1 and occlusal bite block electronics 04 Interruption of the 5 V line 05 Interruption of Vref line 06 Interruption of Vout line 07 Interruption of VGND line 08 Interruption of Co...

Страница 157: ...gged in and the bite block plate is deflected all the way down Call service routine S018 9 The data type intended for the next storage function is displayed in selection field 1 The display of patient...

Страница 158: ...st press the Memory key the R key lights up and then the R key The value for the upper value U is accepted L Adjusting and saving the lower value 1 Deflect the bite block plate all the way down up to...

Страница 159: ...e block must be deflected in such a way that the creep speed is effective in the relevant direction observe green arrows The PWM index values are set correctly if the drive runs at approximately the s...

Страница 160: ...10 The PWM index for upward travel U up is displayed in selection field 1 The PWM index for downward travel D down is displayed in selection field 2 The display of patient symbol key 1 G indicates wh...

Страница 161: ...l service routine S021 3 133 An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready to move the diaphragm 2 Touch the T key A The...

Страница 162: ...cking position for the sensor unit 1 Call service routine S021 4 133 An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready to mo...

Страница 163: ...step 10 Self test of the DCS image data path 1 Call service routine S032 10 133 Once the service routine has been selected DCS test is displayed in selection field 1 2 Press the T key The DCS image da...

Страница 164: ...tallation Manual ORTHOPHOS SL 11 12 6 2 S032 Test step 50 Self test of the FP image data path 1 Call service routine S032 50 133 Once the service routine has been selected FP test is displayed in sele...

Страница 165: ...nsmissions and CAN bus commands The image data are analyzed in the memory of the DX11 It is possible to evaluate the LEDs on DX1 image pan during the synchronized readout sequence Image LED V501 Pan L...

Страница 166: ...Replace sensor unit If error messages are present please implement the measures according to the displayed error message Transfer Minimal image data was received by the sensor unit during the self tes...

Страница 167: ...the default IP address default gateway address and default subnet mask S037 3 yes Configuring boot mode DYNAMIC DHCP AutoIP STATIC fixed address S037 4 yes Manual input of static network settings IP...

Страница 168: ...vice routine S037 1 Once the service routine has been selected the IP address of the unit is displayed in selection field 1 see table Default or static or dynamic is displayed in selection field 2 see...

Страница 169: ...r to the restore operation see table The patient symbol key selected in each case lights up 3 To reset the network data press the Memory key E R key lights up followed by the R key F The default netwo...

Страница 170: ...g 1 Call service routine S037 3 Once the service routine has been selected the current boot mode of the unit is displayed in selection field 1 2 Use arrow keys A to select the required boot mode autom...

Страница 171: ...lit The IP address is displayed in selection field 1 or after pressing the T key number pad B1 is selected Patient symbol key 2 B lit The default gateway is displayed in selection field 1 or after pre...

Страница 172: ...field 1 by pressing the patient symbol keys A B or C see table The patient symbol key selected in each case lights up 3 To change the selected parameter first press the T key E 4 Now use the patient...

Страница 173: ...b 5 Use the arrow keys F to select the digit to be changed in selection field 1 Digit No 12 in the example The corresponding patient symbol key D lights up Selection field 2 displays the value of the...

Страница 174: ...e S017 6 Activating deactivating the remote control S017 8 Selecting the kV mA level series S017 13 S017 14 Enable disable the welcome screen or the lines in the welcome screen S017 15 Activate deacti...

Страница 175: ...site The Sirona Web service is loaded The sub menus of the User menu are displayed in the structure tree 5 Click on the Service button in the header line of the main menu The dialog box for entering t...

Страница 176: ...o the service routines that can be called using the control panel see chapter Perform service routines via the control panel 9 Enter the desired parameters in the service menu and confirm the entries...

Страница 177: ...s or remove a cover For measurements in the area of the power supply terminal the unit must be disconnected from the junction box of the building installation before you start to connect test cables D...

Страница 178: ...y should be switched off at the junction box of the building installation when performing measurements in the vicinity of the power supply unit Select the correct current voltage type and adjust the m...

Страница 179: ...lsed radiation of type Mult O Meter 512L or similar Measuring wires Measuring wires with the following properties dielectric strength 1000V CAUTION Only use a battery operated digital multimeter with...

Страница 180: ...t at least 1 minute before starting to check the line voltage Disconnect the unit from the junction box of the building installation 2 Remove the Bottom profile cover see Service Manual 3 CAUTION Only...

Страница 181: ...he Installation Requirements must be installed If the measured voltage drop falls within the permissible tolerance range see the following table finalize the measurement Permissible voltage drop Concl...

Страница 182: ...ready for exposure in SIDEXIS 4 see Operating Instructions 4 CAUTION Activating the release button triggers X rays Start the exposure by pressing the release button Hold down the release button until...

Страница 183: ...he kV mA level 60 kV 8 mA 7 Use the arrow keys A in selection field 2 to select the radiation time 0 5 s Performing measurements Initiate the radiation Hold the release button pressed until the set ra...

Страница 184: ...the unit wait at least 4 minutes LED V500 on the DX6 must no longer be on before removing the cover on the tube assembly 2 Loosen the screw A and remove the lid of the tube assembly cover B 3 Loosen t...

Страница 185: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 185 Sirona Dental Systems GmbH 12 Startup measurements and tests for USA Canada only Installation Manual ORTHOPHOS SL 12 7 Checking the tube current b...

Страница 186: ...DANGER Do not touch any live components Set the main switch A to I see also Operating Instructions 9 Wait for approx 1 minute 10 Press the R key The unit moves to its starting position Performing meas...

Страница 187: ...ng Instructions 2 Loosen the screw and remove the lid of the tube assembly cover 3 DANGER After switching off the unit wait at least 4 minutes before removing the measuring wires or reinserting the ju...

Страница 188: ...localizers Checking the laser light localizers Preparing the test 1 Insert the forehead and temple supports 2 Set the main switch A to I see also Operating Instructions 3 Wait for approx 1 minute 4 T...

Страница 189: ...ight localizer is switched back off again Checking the 3D laser beam 1 Touch the 3D symbol at the top of the touchscreen The 3D program group is selected 2 Select the exposure program Press the arrow...

Страница 190: ...ed 6 Press on the light localizer key again The light localizer is switched back off again Checking the horizontal laser beam ceph FH Tip When checking or adjusting the light localizer you may use a P...

Страница 191: ...rforming the following tests be sure to observe the radiation protection regulations applicable in your country see Operating Instructions DANGER X rays Radiation is signaled by the message X RAY acti...

Страница 192: ...64 95 142 D3632 192 D3632 031 01 02 02 03 2016 13 Unit adjustment and calibration Sirona Dental Systems GmbH Installation Manual ORTHOPHOS SL A...

Страница 193: ...amic Sensor adjustment Diaphragm adjustment Symmetrie adjustment Sensor calibration DCS If a cephalometer is installed Adjustment of Ceph primary diaphragm Adjustment of Ceph secondary diaphragm Adjus...

Страница 194: ...or exposure H4 04 Plug the sensor into the ceph slot incorrect sensor plugged in Status messages Ready for exposure Exposure not possible Please wait Ready for exposure in XXs X RAY Active Exposure is...

Страница 195: ...the unit 13 1 2 1 Opening the calibration menu IT package is installed is installed 1 Start SIDEXIS 4 2 In SIDEXIS 4 call the ORTHOPHOS SL Device administration configuration menu Configuration menu O...

Страница 196: ...ystem date PC in reverse order Example On 25 03 2016 the service password is 3052 Service mode When you open the calibration menu the unit switches from user mode to the PC service mode blocked by the...

Страница 197: ...16 197 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 1 General information on unit adjustment and calibration b 13 1 2 2 Menu structure The menu is...

Страница 198: ...ta entry and display window C Preview image Shows the exposure to be taken in this stage of the adjustment calibration procedure D Tools pictograph Shows which if any test phantom must be used for thi...

Страница 199: ...ifferent areas the menu also contains the following buttons I Fine adjustment Before the fine adjustment and with the check box unchecked a coarse adjustment is performed 203 J Start acquisition Creat...

Страница 200: ...GmbH 13 1 General information on unit adjustment and calibration Installation Manual ORTHOPHOS SL Toolbar functions Button Function Rotate exposure 90 clockwise Reducing size of section Increasing si...

Страница 201: ...orm adjustment and calibration of the unit in the order as described in the section General information on unit adjustment and calibration 193 Structure tree Resetting adjustment calibration and data...

Страница 202: ...are prefixed by status indicators indicating the current status of the corresponding adjustment or calibration operation Adjustment Calibration Not checked Invalid data record or no record present Ad...

Страница 203: ...on adjustment is preset Always try to use precision adjustment first when adjusting the unit In most cases previous coarse adjustment is not necessary A message window indicates that a coarse adjustme...

Страница 204: ...or adjustment can usually be performed directly via precision adjustment Only in exceptional cases e g if one or several needles are completely outside of the image section in an image acquired with t...

Страница 205: ...sually be performed directly via precision adjustment Only in exceptional cases e g if the exposed image area is completely outside the image section in an image acquired with the Fine adjustment sett...

Страница 206: ...s the ceph secondary diaphragm can be performed directly via precision adjustment Only in exceptional cases e g if the exposed image area is completely outside the image section in an image acquired w...

Страница 207: ...f the exposed image section in the direction of the stationary auxiliary lines see the example Example In the example the exposed image section is offset to the left by the value Sx and upward by the...

Страница 208: ...matic adjustment until the default values in the text boxes are equal to zero This means that adjustment has been successful If the values do not converge toward zero despite repeating the adjustment...

Страница 209: ...on the control panel along with the specific exposure parameters 13 1 4 Taking an exposure The calibration menu must be opened Select the corresponding element in the structure tree The unit must be...

Страница 210: ...justments the needle phantom A must be inserted into the pan bite block holder of the unit For Diaphragm adjustment panorama adjustment the needle phantom must be taken out of the bite block holder NO...

Страница 211: ...nt and calibration Installation Manual ORTHOPHOS SL 13 1 General information on unit adjustment and calibration b When fitting the needle phantom make sure that it is correctly oriented For the panora...

Страница 212: ...OPHOS SL 13 1 6 2 Adjustment phantom for adjustment of the cephalometer For Adjustment of Ceph primary diaphragm and Adjustment of Ceph main beam direction ceph adjustments the ceph adjustment phantom...

Страница 213: ...ometry phantom A into the pan bite block holder B on the unit For Diaphragm adjustment 3D adjustment as well as for Sensor calibration DCS and Diaphragm calibration 3D calibrations you must take the g...

Страница 214: ...Installation Manual ORTHOPHOS SL 1 Insert the geometry phantom A into the pan bite block holder B on the unit and align the geometry phantom correctly with the screw C 2 Attach the clip D for better s...

Страница 215: ...t and calibration via the calibration menu Only for Germany 13 2 1 2D adjustment calibration 13 2 1 1 Pan sensor adjustment IMPORTANT After every adjustment and calibration of the unit the reference v...

Страница 216: ...ure tree click on the Sensor adjustment element S010 1 under Pan The Sensor adjustment menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjus...

Страница 217: ...ting at Step 3 until the values in the menu s text boxes are equal to zero This means that adjustment was performed successfully IMPORTANT If automatic adjustment fails to produce values equal to zero...

Страница 218: ...ibration menu Installation Manual ORTHOPHOS SL 6 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Sensor adjustm...

Страница 219: ...ed in the chapter entitled Automatic adjustment Panoramic sensor standard 216 2 After taking the exposure measure the distances S1 S2 and S3 with the toolbar measuring ruler 197 To measure S1 S2 and S...

Страница 220: ...ustment values for S1 S2 and S3 in the text boxes of the Sensor adjustment menu with the measured values For information about the shifting direction input of sign in the menu see the chapter entitled...

Страница 221: ...the diaphragm is tilted appears in the menu during Diaphragm adjustment the primary diaphragm is mechanically maladjusted in the vertical axis In this case adjust the diaphragm 237 and repeat after t...

Страница 222: ...ement S030 2 The Diaphragm adjustment menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjustment is pre selected Perform a precision adjustm...

Страница 223: ...64 95 142 D3632 D3632 031 01 02 02 03 2016 223 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b...

Страница 224: ...zero calculate the adjustment values manually and overwrite the values in the menu s text boxes with these values 225 5 If adjustment is OK the adjustment values in the text boxes are equal to zero s...

Страница 225: ...adjustment The only difference is that the default adjustment values calculated automatically in the Diaphragm adjustment menu are overwritten by adjustment values calculated manually 1 Start the Diap...

Страница 226: ...justment values for Sx and Sy in the text boxes of the Diaphragm adjustment menu with the measured values For information about the shifting direction input of sign in the menu see the chapter entitle...

Страница 227: ...mmetry The needle phantom is in the bite block holder on the unit 210 1 In the structure tree under Pan click on the Symmetrie adjustment element S010 2 The Pan Symmetrie menu is displayed in the acti...

Страница 228: ...64 95 142 D3632 228 D3632 031 01 02 02 03 2016 13 Unit adjustment and calibration Sirona Dental Systems GmbH 13 2 Adjustment and calibration via the calibration menu Installation Manual ORTHOPHOS SL...

Страница 229: ...u s text boxes with these values 230 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Symmetrie adjustment ele...

Страница 230: ...pter entitled Automatic adjustment Pan Symmetry 227 2 After taking the exposure measure the distances S1 S2 and S3 with the toolbar measuring ruler 197 If possible measure in the lower area of the nee...

Страница 231: ...n the structure tree under Pan click on the Sensor calibration DCS element S010 10 3 Establish 209 receptivity 4 Take the exposure for the first kV mA combination 60kV 5mA 15 1s 209 After taking the e...

Страница 232: ...ombinations After the last exposure has been taken a message in the action window indicates whether calibration has been successful and the status indicator at the start of the element Sensor calibrat...

Страница 233: ...m connects to the cover of the sensor unit 212 The ceph sensor is plugged into the ceph slot on the unit The ear plug holders on the ceph arm have been moved completely apart and swung out of the beam...

Страница 234: ...d calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 64 kV 16 mA 6 1 s 209 The adjustment values S1 S2 S3 and S4 are automatically determined from the exposure...

Страница 235: ...3632 D3632 031 01 02 02 03 2016 235 Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b A B S2 S2...

Страница 236: ...menu s text boxes with these values 237 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 The status indicator that prefixes the Adjustment of Ceph pr...

Страница 237: ...in the chapter entitled Automatic adjustment Ceph Primary diaphragm standard 233 2 Once you have taken the exposure measure distances S1 S2 S3 and S4 with the toolbar measuring ruler 197 S1 distance...

Страница 238: ...in the pan bite block holder on the unit The ceph test phantom does not connect to the cover of the sensor unit IMPORTANT If you cannot correctly adjust the ceph secondary diaphragm via the calibrati...

Страница 239: ...diaphragm menu is displayed in the action area The menu supports precision adjustment and coarse adjustment precision adjustment is pre selected Perform a precision adjustment first In most cases prev...

Страница 240: ...tion via the calibration menu Installation Manual ORTHOPHOS SL A Adjustment OK The exposed diaphragm area must lie centered and straight in the image section as well as inside the superimposed auxilia...

Страница 241: ...o This means that adjustment was performed successfully IMPORTANT If automatic adjustment fails to produce values equal to zero calculate the adjustment values manually and overwrite the values in the...

Страница 242: ...hragm The manual adjustment procedure is similar to the one for automatic adjustment 238 The only difference is that the default adjustment values calculated automatically in the Adjustment of Ceph se...

Страница 243: ...ntal Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 2 Once you have taken the exposure measure distances S1...

Страница 244: ...for S1 S2 and S3 in the text boxes of the Adjustment of Ceph secondary diaphragm menu with the measured values For information about the shifting direction input of sign in the menu see the section en...

Страница 245: ...d calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 13 2 1 6 3 Mechanical adjustment Ceph secondary diaphragm 1 Loosen screw A and remove the cove...

Страница 246: ...rn counterclockwise adjusts the diaphragm to the right Turn clockwise adjusts the diaphragm to the left Screw D Turn counterclockwise adjusts the diaphragm downward Turn clockwise adjusts the diaphrag...

Страница 247: ...an needle phantom is not in the pan bite block holder on the unit The ceph test phantom does not connect to the cover of the sensor unit 1 In the structure tree under Ceph click on the Adjustment of C...

Страница 248: ...nt and calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 80kV 14 mA 0 3 s 209 Adjustment value S1 is calculated automatically from the exposure and entered in...

Страница 249: ...djustment values manually and overwrite the values in the menu s text boxes with these values 242 5 If adjustment is OK the adjustment values in the text boxes are equal to zero save the values 209 Th...

Страница 250: ...nual adjustment procedure is similar to the one for automatic adjustment 247 The only difference is that the default adjustment values calculated automatically in the Adjustment of Ceph secondary diap...

Страница 251: ...Sirona Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 2 Once you have taken the exposure measure di...

Страница 252: ...S1 in the text box of the Adjustment of Ceph secondary diaphragm Quickshot menu with the measured value For information about the shifting direction input of sign in the menu see the chapter entitled...

Страница 253: ...connects to the unit s sensor unit 212 The ear plug holders on the ceph arm have been swung out of the beam direction ap 1 In the structure tree under Ceph click on the Adjustment of Ceph main beam di...

Страница 254: ...al Systems GmbH 13 2 Adjustment and calibration via the calibration menu Installation Manual ORTHOPHOS SL 3 Create an exposure 80kV 14mA 14 9 s 209 Adjustment value S1 is calculated automatically from...

Страница 255: ...es equal to zero calculate the adjustment values manually and overwrite the values in the menu s text boxes with these values 256 5 If adjustment is OK the adjustment values in the text boxes are equa...

Страница 256: ...matic adjustment Ceph Main X ray beam direction standard 253 2 Once you have taken the exposure measure the distance S1 with the toolbar measuring ruler 197 Measure within the measuring range shown in...

Страница 257: ...earplug alignment 13 2 1 9 1 Checking the ear plug alignment The ceph sensor is plugged into the ceph slot on the unit 1 In the structure tree under Ceph click on the Adjustment of earplug alignment...

Страница 258: ...ide sensor side Transparent adjusting cap B on the inside tube assembly side 4 Establish 209 receptivity 5 Create an exposure 80kV 14mA 14 9 s 209 The exposure is displayed in the exposure window The...

Страница 259: ...na Dental Systems GmbH 13 Unit adjustment and calibration Installation Manual ORTHOPHOS SL 13 2 Adjustment and calibration via the calibration menu b 13 2 1 9 2 Adjusting the ear plug alignment Unscre...

Страница 260: ...o not unscrew them fully 2 Adjust the ear plugs in the horizontal direction by turning the screw B counterclockwise or clockwise The direction of rotation for left arm C and right arm D versions is sh...

Страница 261: ...ibration menu b 3 Adjust the ear plugs in the vertical direction by turning the knurled nut F The direction of rotation for left arm G and right arm H versions is shown in the drawing below 4 Tighten...

Страница 262: ...2 3D adjustment calibration 13 2 2 1 Sensor adjustment 1 Insert the 3D calibration block into the bite block holder on the unit 213 IMPORTANT For the 3D adjustment and calibration the values in the in...

Страница 263: ...element S010 20 5 Establish readiness for exposure 209 6 Create an exposure 85kV 22 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successful and the value...

Страница 264: ...S030 10 2 Establish readiness for exposure 209 3 Create an exposure 85kV 0 34 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successful and the values can...

Страница 265: ...tion element S010 64 2 Establish 209 receptivity 3 Create an exposure 85kV 22 mAs 209 A message is displayed in the action window to indicate that the exposure is complete and the status indicator in...

Страница 266: ...ment S030 11 The 5x5 volume is selected 2 Establish 209 receptivity 3 Create an exposure 85kV 0 34 mAs 209 A message is displayed in the action window to indicate whether adjustment has been successfu...

Страница 267: ...status indicator turns green NOTE The calibration values are automatically saved 7 If volume 11x10 is available Click on the Next button or select the volume 11x10 in the structure tree S030 13 The 1...

Страница 268: ...tree under 3D Geometry calibration 5x5 click on the Molar left element S010 25 S010 39 The Molar left element is selected 4 Establish 209 receptivity 5 Create an exposure 85kV 22 mAs 209 A message is...

Страница 269: ...been successful and the status indicator in front of the 5x5 Premolar left element in the structure tree responds accordingly Calibration not OK status indicator stays red Calibration OK status indic...

Страница 270: ...21 2 Attach the Mult O Meter sensor in the middle of the sensor unit flat panel detector side 3 Select a program from the 3D program group on the control panel The unit moves to the 3D exposure positi...

Страница 271: ...the structure tree under Reset on the element Reset Pan Reset Ceph or Reset 3D depending on what data you would like to reset 2 To start resetting the adjustment and calibration data click on the Res...

Страница 272: ...lorer or Firefox 2 In the address line enter https and the IP address of your unit Example https 192 168 15 240 Tip The IP address of your unit can be found on the info screen 3 Confirm your input wit...

Страница 273: ...ting the touchscreen via the web service b A message is displayed in the web browser to indicate that the adjustment screen has been opened on the display The first adjustment screen appears on the co...

Страница 274: ...ms GmbH 13 3 Adjusting the touchscreen via the web service Installation Manual ORTHOPHOS SL 8 Touch the center of the square in the bottom right corner of the screen 9 Touch the center of the square i...

Страница 275: ...nt and calibration Installation Manual ORTHOPHOS SL 13 3 Adjusting the touchscreen via the web service b The unit switches to PC service mode Service mode is displayed on the Easypad via the PC servic...

Страница 276: ...oard following replacement of the board This is possible via the Restore sensor data menu see the Service Manual Open the Backup Restore sensor data Open the backup restore data menu The unit and comp...

Страница 277: ...4 1 Save sensor data b 5 Click on the Service button in the header line of the main menu The dialog box for entering the access data opens 6 Enter the following access data Username service Password s...

Страница 278: ...or data element The Backup sensor data menu is displayed 2 Click on the Backup sensor data button The Download Backup menu opens A selection dialog box opens in your browser 3 In the selection dialog...

Страница 279: ...Opening Unit information b 14 2 Opening Unit information SIDEXIS 4 is installed The workstation software for the corresponding X ray component or sensor is installed on the workstation 1 Start SIDEXI...

Страница 280: ...016 14 Final work Sirona Dental Systems GmbH 14 2 Opening Unit information Installation Manual ORTHOPHOS SL The Device configuration menu opens 4 Click on the Device details button in the structure tr...

Страница 281: ...ona Dental Systems GmbH 14 Final work Installation Manual ORTHOPHOS SL 14 2 Opening Unit information b The Unit information Advanced information is displayed 5 Click on the Downloading button to load...

Страница 282: ...Operating system according to the op systems approved in the installation instructions Removable disk drive1 DVD ROM DVD RAM 1 supported by SIDEXIS function required only if not performed by an additi...

Страница 283: ...Installation Manual ORTHOPHOS SL 14 4 Unit handover b 14 4 Unit handover Unit handover Give the customer all of the patient positioning aids test phantoms and special tools included in delivery inclu...

Страница 284: ...t K p a p d e OMNS p W m d aPSPOKMPNKMNKMOKMO MPKOMNS KJk KW NON UNO p a p d e rp W c PN aJSQSOR _ d p a I f K QUPR p a I k OUOTP rp l k SQ VR NQO aPSPO K K...

Отзывы: