64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
221
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.2
Pan aperture adjustment
13.2.1.2.1
Automatic adjustment: Pan diaphragm (standard)
IMPORTANT
If a message window indicating that the diaphragm is tilted appears in
the menu during
"Diaphragm adjustment"
, the primary diaphragm is
mechanically maladjusted in the vertical axis. In this case, adjust the
diaphragm [ → 237] and repeat after the exposure.
IMPORTANT
If a message window indicating that another X-ray has to be taken
appears in the menu during
"Diaphragm adjustment"
, repeat the
procedure starting at Step 2, even if the values in the menu's text boxes
are equal to zero. This additional exposure ensures that the diaphragm
gap width offset is also adjusted correctly.