64 95 142 D3632
174
D3632
.
031.01.02
.
02 03.2016
11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.8
Running service routines via the web service
NOTE: The web service is only available in English
NOTE: The following service routines can be run using the Sirona Web
service:
In the main menu
"Service"
● S008.2: Overview of module software versions
● S008.3: Confirming the unit serial number
● S017.2: Configuring the hardware version
● S017.3: Enter the country group code
● S017.4: Select a language
● S017.6: Activating / deactivating the remote control
● S017.8: Selecting the kV/mA level series
● S017.13/S017.14: Enable/disable the welcome screen or the lines in
the welcome screen
● S017.15: Activate/deactivate
the acoustic signal for end of exposure
● S017.18: Activation/deactivation of occlusal bite block function
● S037: Manual input of static network settings (IP address, default
gateway address, and subnet mask)
✔
The unit and computer are turned on.
✔
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
NOTICE! The IP address of your unit can be found on the Info screen.
3. Confirm your input with the "Enter" key.
A security prompt appears.