11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.5.2
S021: Test step 4
Initiate the packing position for the sensor unit
1. Call service routine S021.4 [ → 133].
An inactive progress indicator in selection field 1 and the
message FFFF in selection field 2 signal that the system is ready
to move the sensor unit.
2. Touch the T key (A).
The sensor unit is moved to the packing position.
During this time, a progress indicator is displayed.
When the packing position is reached, "0000" is displayed in
selection field 2.