11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.3.5
S017: Test step 8
Selecting the kV/mA level series
* Factory setting
1. Call service routine S017.8.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
Selection field
Code
Function
1
1E
16 mA series for pan
16 mA series for Ceph
2E
8 mA series for pan
16 mA series for Ceph