Using the manual block clamp
Sirona Dental Systems GmbH
Installation Instructions for Manual Block Clamp CEREC MC XL, inLab MC XL
5
Using the manual block clamp
Store the block clamp tool in the corresponding holder.
You can attach the holder to a suitable location with the adhesive pad. Clean
and degrease the contact surface beforehand.
empty
Blocks with 10 mm diameter block holders
Fastening the block
Instructions for 10 mm diameter
1. Place the block (A) directly into the block clamp.
2. Clamp the ceramic block securely with the ball pressure screw (B). Use
the block clamp tool with torque wrench for this purpose.
Ç
The block is pressed laterally against the contact surface of the block
clamp and simultaneously pulled in axially. The plate of the block hol
-
der thus rests on the block clamp.
NOTICE
Wear of the ball pressure screw
The high clamping forces cause wear of the ball pressure screw.
Replace the ball pressure screw every 500 clamping operations.
NOTICE
Fasten the block tightly
If the block is not tightened sufficiently, this may result in falsification of the
milling result and fracturing of its ceramic material.
Fasten the block tightly using the block clamp tool with torque wrench un
-
til you hear a click.
Check to make sure that the block is seated very firmly.
Содержание inLab MC XL
Страница 2: ......
Страница 3: ...MQKOMNM bob j uiI áåi Ä j ui kÉì ÄW aÉìíëÅÜ báåÄ ì åäÉáíìåÖ ÑΩê ã åìÉääÉ _äçÅâëé ååìåÖ ...
Страница 4: ......
Страница 19: ...MQKOMNM bob j uiI áåi Ä j ui kÉï ë çÑW båÖäáëÜ fåëí ää íáçå fåëíêìÅíáçåë Ñçê j åì ä _äçÅâ ä ãé ...
Страница 35: ...MQKOMNM bob j uiI áåi Ä j ui kçìîÉ ì ÇÉéìáëW cê å áë fåëíêìÅíáçåë ÇÉ ãçåí ÖÉ Çì ëÉêêÉJÄäçÅ ã åìÉä ...
Страница 51: ...MQKOMNM bob j uiI áåi Ä j ui kìÉîç é êíáê ÇÉW bëé çä fåëíêìÅÅáçåÉë ÇÉ ãçåí àÉ ÇÉä ëìàÉí ÄäçèìÉ ã åì ä ...
Страница 83: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ...NOTE BELOW THE LAST POSITION WHEN BE FULL AND RIGHT INSTALLED ...