62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
71
Sirona Dental Systems GmbH
Appendix
Installation Instructions HELIODENT
PLUS
Replacing old units
båÖäáëÜ
9.3.2
Replacement of horizontally mounted old units with an
adapter plate
empty
Explanation
The adapter plate (REF 62 42 254) serves to conceal an existing
installation site of a horizontally mounted old unit and provide the
threaded holes required to mount a HELIODENT
PLUS
.
empty
Suitable old units
● HELIODENT DS
● HELIODENT MD
● Planmeca Intra
empty
Prerequisites
● A check was performed to ensure that the mounting dimensions of
the old unit match those of the adapter plate.
empty
CAUTION
The on-site electrical installation must be performed according to the
valid regulations for medical electrical equipment (DIN VDE 0100-710).
CAUTION
Observe wall properties
In installation situations, the technician is responsible for the
assessment of wall properties and selecting the method of attaching the
unit to the wall.
NOTICE
Exclusion of liability if the adapter plate is modified
If any design/structural changes are made in the adapter plate (e.g.
additional holes, removal of material, etc.) Sirona accepts no liability
whatsoever for any damage caused by said changes.