61 25 665 D3437
364
D3437
.
076.01.15
.
02 06.2012
9 Repair
Sirona Dental Systems GmbH
9.13 Cable
Service Manual GALILEOS
9.13.2.2
Cable replacement (L3, L5, L6, and L15)/Laying the cable/
corrugated tube at the rotation unit
Cable replacement (L3, L5, L6, and L15)/Laying the cable/corrugated tube at the rotation unit
Prepare the cable exchange
1. Remove the covers.
2. Pull the connectors off board DX6.
Replacing cables
1. For L3: Remove the connector of the cable with tool W1.
2. Remove the corrugated tube and the spiral spring (B) from the cable
loom.
3. Remove the defective cable and run the new cable up to the rotary
ring in the original position.
4. Bunch the cables together again to form a loom.
5. Fasten the defective cable to the loom and use it as a pull wire to pull
the loom through the fabric tube (A).
6. Pull the fabric tube over the connector and as far over the cable loom
as possible.
7. Use the pull wire to pull the fabric tube into the spiral spring (B).
NOTICE
The connectors and cables must be protected by inserting them in the
fabric tube (A) supplied with the cables.