61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
335
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.9
Facescan
båÖäáëÜ
9.9.2.2
Installing the new PoE module
Installing the new PoE module
1. Put the PoE module (P) on the thread bolts (H) of the mains filter plate
(J) (the serrated washer must sit behind the fitting plate of the PoE
module).
2. Screw down the PoE module (P) to the place provided in the stand
with two screws (K) and a nut (L).
3. Plug Ethernet cable L77 into the POE socket of the PoE module.
4. To fasten, slide the cable cover (I) downwards again as far as
possible.
IMPORTANT
The locking catch of the RJ45 plug of Ethernet cable L77 is secured with
adhesive tape (N).
➢ Remove the adhesive tape (N) from the RJ45 plug of Ethernet cable
L77.
Содержание galileos
Страница 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Страница 2: ......
Страница 387: ......