61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
325
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.9
Facescan
båÖäáëÜ
9.9
Facescan
Facescan
9.9.1
Replacing the scan unit
Replacing the scan unit
9.9.1.1
Removing the defective scan unit
Removing the defective scan unit
Removing the covers
1. Undo the four screws (K) and remove the two covers (J).
2. Disconnect the cable (H) from the display board (I) on the inside of
the cover (F).
3. Undo the six screws (G) and remove the cover (F).