10 Checking the safety switches
Sirona Dental Systems GmbH
9.3 Care and cleaning of the monitor
59 06 511 D 3386
28
D 3386.103.01.06.02
10
Checking the safety switches
To ensure the safety of the practice team and the
patients, please check the safety switches installed in
the unit every 6 months.
Move the chair or the backrest down for this check.
• Raise the leg support:
The chair movement stops, then there is a corrective
movement upwards for around 3sec.
• Raise the backrest:
The backrest movement stops, then there is a correc-
tive movement upwards for around 3sec.
• Raise the motor-adjustable headrest:
The backrest movement stops, then there is a correc-
tive movement upwards for around 3sec.
• Swing in the swiveling cuspidor (optional):
The chair movement stops.
ATTENTION
If one of the safety switches does not stop the chair
movement or the backrest movement, the treatment
center must not be used for dental work!
Please contact your service engineer.
Содержание C1+
Страница 2: ......
Страница 4: ......
Страница 5: ... NH mÑäÉÖÉ ìåÇ oÉáåáÖìåÖ ÇìêÅÜ Ç ë mê ñáëíÉ ã aÉìíëÅÜ ...
Страница 8: ...Inhaltsverzeichnis Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 01 ...
Страница 33: ......
Страница 35: ... NH êÉ åÇ äÉ åáåÖ Äó íÜÉ mê ÅíáÅÉ qÉ ã båÖäáëÜ ...
Страница 38: ...Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 02 ...
Страница 63: ......
Страница 65: ... NH båíêÉíáÉå Éí åÉííçó ÖÉ é ê äÛ èìáéÉ Çì Å ÄáåÉí cê å áë ...
Страница 68: ...Sommaire Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 03 ...
Страница 93: ......
Страница 95: ... NH ìáÇ Çç ó äáãéáÉò éçê é êíÉ ÇÉä éÉêëçå ä ÇÉ ä Åçåëìäí bëé çä ...
Страница 98: ...IÍndice Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 04 ...
Страница 123: ......
Страница 125: ......